Device RecallsRecall Watch

Recall Watch · FDA Class II

FDA Class II Recall Watch: 20–26 September 2026

Eight FDA Class II recall events cover pregnancy tests, respiratory and imaging systems, biopsy equipment, surgical kits, embolization devices, and bone cement. Corrective actions include component replacement, software updates, inventory controls, and product returns.

Published Issue FDA posting period 20–26 September 2026

At a glance

A compact index of this issue. Select an entry to move to its complete record.

Selected FDA Class II recall entries in this issue
Company/ProductRecall issueOfficial actionQuantityEntry link
Alere San Diego, Inc.Alere hCG rapid pregnancy testsAlere recalls multiple hCG rapid-test configurations over potential false-positive resultsFor unexpected results, customers were instructed to consider the test limitations, including low hCG levels associated with peri-menopause, and to consider an alternate confirmatory test when making a pregnancy diagnosis.Z-3271-2026: 50,265,835; Z-3272-2026: 5,331,585; Z-3273-2026: 13,256,075; Z-3274-2026: 18,139,960Alere recalls multiple hCG rapid-test configurations over potential false-positive results — read full entry
NOXBOX LTDNOxBOXi Nitric Oxide Delivery SystemNOXBOX recalls NOxBOXi systems over potential irregular oxygen readingsDistributors and device users were instructed to examine inventory for affected units and return the applicable response form.1,667 unitsNOXBOX recalls NOxBOXi systems over potential irregular oxygen readings — read full entry
Volcano CorpPhilips IntraSight PlusVolcano recalls Philips IntraSight Plus software over patient-image association riskCustomers were instructed to contact Philips service before changes to the worklist provider, Azurion software, interface configuration, or hospital IT environment.19 systemsVolcano recalls Philips IntraSight Plus software over patient-image association risk — read full entry
Bard Peripheral Vascular IncEnCor EnCompass Breast Biopsy and Tissue Removal SystemBard recalls EnCor EnCompass systems over biopsy-procedure software errorsIf probe calibration cannot be completed after troubleshooting, clinicians were instructed to replace the probe and, if necessary, use an alternative biopsy driver or system.99Bard recalls EnCor EnCompass systems over biopsy-procedure software errors — read full entry
Merit Medical Systems, Inc.Embosphere MicrospheresMerit Medical recalls Embosphere Microspheres over sterile-barrier packaging defectsCustomers were instructed to discontinue use and return affected devices to Merit Medical.104Merit Medical recalls Embosphere Microspheres over sterile-barrier packaging defects — read full entry
Medline Industries, LPMedline convenience kits with neurological spongesMedline recalls convenience kits containing neurological sponges over endotoxin variabilityCustomers were instructed to identify and quarantine affected kits and submit the recall response form, including affected inventory quantities.5,556 kits across five FDA recall recordsMedline recalls convenience kits containing neurological sponges over endotoxin variability — read full entry
AVID Medical, Inc.Halyard IV START KIT NON-STERILEAVID Medical recalls Halyard IV START kits containing affected BD ChloraPrep applicatorsFor kits containing applicators affected by the contamination update, consignees were instructed to discard the impacted kits and notify customers that received them.1,008 kits (21 cases) across two FDA recall recordsAVID Medical recalls Halyard IV START kits containing affected BD ChloraPrep applicators — read full entry
Heraeus Medical GmbH (Dental Division)PALACOS pro bone-cement systemsHeraeus recalls PALACOS pro bone-cement systems over potential monomer leakageConsignees were instructed to ensure relevant hospital departments received the correction letter and to wear suitable eye protection whenever the system is operated.Z-3301-2026: 33,727 units; Z-3302-2026: 59,660 units; Z-3303-2026: 20,523 units; Z-3304-2026: quantity not reportedHeraeus recalls PALACOS pro bone-cement systems over potential monomer leakage — read full entry

Entry 01

Alere recalls multiple hCG rapid-test configurations over potential false-positive results

Company/Product
Alere San Diego, Inc. — Alere hCG rapid pregnancy tests
Recall issue
Alere recalls multiple hCG rapid-test configurations over potential false-positive results
Official action
For unexpected results, customers were instructed to consider the test limitations, including low hCG levels associated with peri-menopause, and to consider an alternate confirmatory test when making a pregnancy diagnosis.
Quantity
Z-3271-2026: 50,265,835; Z-3272-2026: 5,331,585; Z-3273-2026: 13,256,075; Z-3274-2026: 18,139,960
Alere recalls multiple hCG rapid-test configurations over potential false-positive results — read full entry

Entry 02

NOXBOX recalls NOxBOXi systems over potential irregular oxygen readings

Company/Product
NOXBOX LTD — NOxBOXi Nitric Oxide Delivery System
Recall issue
NOXBOX recalls NOxBOXi systems over potential irregular oxygen readings
Official action
Distributors and device users were instructed to examine inventory for affected units and return the applicable response form.
Quantity
1,667 units
NOXBOX recalls NOxBOXi systems over potential irregular oxygen readings — read full entry

Entry 03

Volcano recalls Philips IntraSight Plus software over patient-image association risk

Company/Product
Volcano Corp — Philips IntraSight Plus
Recall issue
Volcano recalls Philips IntraSight Plus software over patient-image association risk
Official action
Customers were instructed to contact Philips service before changes to the worklist provider, Azurion software, interface configuration, or hospital IT environment.
Quantity
19 systems
Volcano recalls Philips IntraSight Plus software over patient-image association risk — read full entry

Entry 04

Bard recalls EnCor EnCompass systems over biopsy-procedure software errors

Company/Product
Bard Peripheral Vascular Inc — EnCor EnCompass Breast Biopsy and Tissue Removal System
Recall issue
Bard recalls EnCor EnCompass systems over biopsy-procedure software errors
Official action
If probe calibration cannot be completed after troubleshooting, clinicians were instructed to replace the probe and, if necessary, use an alternative biopsy driver or system.
Quantity
99
Bard recalls EnCor EnCompass systems over biopsy-procedure software errors — read full entry

Entry 06

Medline recalls convenience kits containing neurological sponges over endotoxin variability

Company/Product
Medline Industries, LP — Medline convenience kits with neurological sponges
Recall issue
Medline recalls convenience kits containing neurological sponges over endotoxin variability
Official action
Customers were instructed to identify and quarantine affected kits and submit the recall response form, including affected inventory quantities.
Quantity
5,556 kits across five FDA recall records
Medline recalls convenience kits containing neurological sponges over endotoxin variability — read full entry

Entry 07

AVID Medical recalls Halyard IV START kits containing affected BD ChloraPrep applicators

Company/Product
AVID Medical, Inc. — Halyard IV START KIT NON-STERILE
Recall issue
AVID Medical recalls Halyard IV START kits containing affected BD ChloraPrep applicators
Official action
For kits containing applicators affected by the contamination update, consignees were instructed to discard the impacted kits and notify customers that received them.
Quantity
1,008 kits (21 cases) across two FDA recall records
AVID Medical recalls Halyard IV START kits containing affected BD ChloraPrep applicators — read full entry

Entry 08

Heraeus recalls PALACOS pro bone-cement systems over potential monomer leakage

Company/Product
Heraeus Medical GmbH (Dental Division) — PALACOS pro bone-cement systems
Recall issue
Heraeus recalls PALACOS pro bone-cement systems over potential monomer leakage
Official action
Consignees were instructed to ensure relevant hospital departments received the correction letter and to wear suitable eye protection whenever the system is operated.
Quantity
Z-3301-2026: 33,727 units; Z-3302-2026: 59,660 units; Z-3303-2026: 20,523 units; Z-3304-2026: quantity not reported
Heraeus recalls PALACOS pro bone-cement systems over potential monomer leakage — read full entry

This week’s entries

Official recall facts and actions are separated from MedTech Current significance analysis.

Entry 01FDA Class II

Alere recalls multiple hCG rapid-test configurations over potential false-positive results

Multiple Alere-manufactured urine and serum hCG rapid pregnancy test cassettes, strips, dipsticks, and combo tests distributed under Alere, McKesson, Fisher Scientific, Cardinal Health, Medline, Henry Schein, Beckman Coulter, and Pro Advantage brands

Status
Open, Classified
Recalling firm
Alere San Diego, Inc.
Manufacturer
Alere San Diego, Inc.
Event ID
99409
Recall number
Z-3271-2026, Z-3272-2026, Z-3273-2026, Z-3274-2026
FDA posting date
Recall initiated
Quantity
Z-3271-2026: 50,265,835; Z-3272-2026: 5,331,585; Z-3273-2026: 13,256,075; Z-3274-2026: 18,139,960
Distribution
Worldwide distribution, including U.S. nationwide distribution and listed international markets.

Official recall information

What happened

FDA posted four Class II recall records covering multiple urine and serum hCG rapid pregnancy-test configurations made by Alere San Diego. The tests may detect hCG concentrations below their stated assay sensitivity, which may produce false-positive results. FDA lists separate quantities of 50,265,835; 5,331,585; 13,256,075; and 18,139,960 across the four records, without specifying a quantity unit.

MedTech Current analysis

Why it matters

False-positive pregnancy results may affect subsequent clinical decisions. Because the affected tests appear under several brands, laboratories and distributors need to trace which configurations they supplied and ensure downstream users receive the correction notice.

Official recall information

Official actions

According to the FDA recall records:

  • For unexpected results, customers were instructed to consider the test limitations, including low hCG levels associated with peri-menopause, and to consider an alternate confirmatory test when making a pregnancy diagnosis.
  • Customers that forwarded affected products to other laboratories were instructed to notify those laboratories and provide the field-corrective-action letter.
  • Recipients were instructed to retain the letter and acknowledge receipt.
Entry 02FDA Class II

NOXBOX recalls NOxBOXi systems over potential irregular oxygen readings

NOxBOXi Nitric Oxide Delivery System, model NOXBOX-I

Status
Open, Classified
Recalling firm
NOXBOX LTD
Manufacturer
NOXBOX LTD
Event ID
99816
Recall number
Z-3306-2026
FDA posting date
Recall initiated
Quantity
1,667 units
Distribution
US Nationwide distribution in the state of Georgia.

Official recall information

What happened

FDA classified the recall of specified NOxBOXi Nitric Oxide Delivery Systems as Class II because certain oxygen sensors may produce irregular oxygen readings due to a manufacturing issue. The affected devices are model NOXBOX-I and FDA lists 1,667 units in commerce. The recall covers specified serial numbers, with the FDA record listing U.S. distribution in Georgia.

MedTech Current analysis

Why it matters

Oxygen readings guide respiratory care, so affected sensors need to be identified before replacement. Clinical engineering and respiratory teams also need to coordinate with distributors on response forms and the timing of sensor changes.

Official recall information

Official actions

According to the FDA recall record:

  • Distributors and device users were instructed to examine inventory for affected units and return the applicable response form.
  • The firm instructed distributors to arrange replacement of affected oxygen sensors and perform preventive-maintenance checks before distributing units.
  • Organizations were instructed to notify applicable personnel and customers that received affected units.
Entry 03FDA Class II

Volcano recalls Philips IntraSight Plus software over patient-image association risk

Philips IntraSight Plus systems with software version 6.0.4

Status
Open, Classified
Recalling firm
Volcano Corp
Manufacturer
Volcano Corp
Event ID
99730
Recall number
Z-3276-2026
FDA posting date
Recall initiated
Quantity
19 systems
Distribution
Worldwide distribution, including U.S. distribution in MA, MN, TX, AL, GA, and CA.

Official recall information

What happened

FDA classified the correction of Philips IntraSight Plus systems running software version 6.0.4 as Class II. The software may associate image data from one patient with another patient's record, potentially contributing to incorrect diagnosis or treatment. FDA lists 19 systems in commerce, with distribution in six U.S. states and several European countries.

MedTech Current analysis

Why it matters

Images associated with the wrong patient can mislead clinical interpretation even when the image itself is valid. Imaging and hospital IT teams should account for the Study ID risk before changing worklist, interface, or system configurations.

Official recall information

Official actions

According to the FDA recall record:

  • Customers were instructed to contact Philips service before changes to the worklist provider, Azurion software, interface configuration, or hospital IT environment.
  • The notice was to be circulated to IntraSight Plus users and posted on or near the system until Philips updates the software.
  • The FDA record states that the device may otherwise continue to be used according to its intended use.
Entry 04FDA Class II

Bard recalls EnCor EnCompass systems over biopsy-procedure software errors

EnCor EnCompass Breast Biopsy and Tissue Removal System Console and Driver with software versions 1.1.1.10207 and 1.2.0.12133

Status
Open, Classified
Recalling firm
Bard Peripheral Vascular Inc
Manufacturer
Bard Peripheral Vascular Inc
Event ID
99590
Recall number
Z-3275-2026
FDA posting date
Recall initiated
Quantity
99
Distribution
US Nationwide distribution in listed states.

Official recall information

What happened

FDA classified the EnCor EnCompass Breast Biopsy and Tissue Removal System correction as Class II. Software errors may cause probe-calibration, handheld-driver, or pressure-sensor failures, interrupting or prolonging a biopsy and potentially requiring additional sampling or a repeat procedure. FDA lists 99 affected products in commerce across multiple U.S. states.

MedTech Current analysis

Why it matters

An interruption during breast biopsy may delay sampling or require a repeat procedure. Facilities need a workable fallback for affected systems and access for the planned software update, with clinicians informed before the next use.

Official recall information

Official actions

According to the FDA recall record:

  • If probe calibration cannot be completed after troubleshooting, clinicians were instructed to replace the probe and, if necessary, use an alternative biopsy driver or system.
  • If a pressure-sensor error cannot be cleared and normal operation restored, users were instructed to discontinue the affected system and complete or reschedule the procedure with alternative equipment.
  • Facilities were instructed to post the correction notice near affected devices, make systems available for the software update, notify relevant users or downstream recipients, and return the response form.
Entry 05FDA Class II

Merit Medical recalls Embosphere Microspheres over sterile-barrier packaging defects

Embosphere Microspheres in 500–700, 700–900, and 900–1200 micrometer 2 mL syringe configurations

Status
Open, Classified
Recalling firm
Merit Medical Systems, Inc.
Manufacturer
Merit Medical Systems, Inc.
Event ID
99541
Recall number
Z-3305-2026
FDA posting date
Recall initiated
Quantity
104
Distribution
Worldwide distribution, including U.S. distribution in CA and GA and distribution to France, Ireland, Canada, and Switzerland.

Official recall information

What happened

FDA classified specified Embosphere Microspheres as a Class II recall because holes may be present in product packaging, potentially compromising the sterile barrier. The microspheres are used for embolization of tumors, prostatic arteries, arteriovenous malformations, and blood vessels; FDA states that use of affected product could lead to infection. The record lists 104 affected products in commerce.

MedTech Current analysis

Why it matters

Holes in the package may compromise sterility before an invasive embolization procedure. Interventional facilities need to locate the affected syringes, stop using them, and trace any stock supplied to other sites.

Official recall information

Official actions

According to the FDA recall record:

  • Customers were instructed to discontinue use and return affected devices to Merit Medical.
  • Organizations were instructed to make applicable personnel aware of the recall and immediately share the notice with facilities or institutions that received further-distributed product.
  • Customers were instructed to complete and return the response form.
Entry 06FDA Class II

Medline recalls convenience kits containing neurological sponges over endotoxin variability

Medline convenience kits classified as ENT, laparoscopy, general surgery, neurological, and orthopedic trays containing affected neurological sponges

Status
Open, Classified
Recalling firm
Medline Industries, LP
Manufacturer
Medline Industries, LP
Event ID
99739
Recall number
Z-3265-2026, Z-3266-2026, Z-3267-2026, Z-3268-2026, Z-3269-2026
FDA posting date
Recall initiated
Quantity
5,556 kits across five FDA recall records
Distribution
US Nationwide.

Official recall information

What happened

FDA posted five Class II recall records for Medline convenience kits containing neurological sponges with potential unexpected variability in endotoxin levels. The records span ENT, laparoscopy, general-surgery, neurological, and orthopedic tray classifications. Their reported quantities are compatible and total 5,556 kits. FDA states that potentially elevated endotoxin may cause neurovascular events, febrile response, inflammation, hypotension, or nausea.

MedTech Current analysis

Why it matters

The same affected sponge component appears in five types of procedure kit. Hospitals therefore need to locate finished kits across specialties and remove the component as directed, while distributors pass the notice to downstream customers.

Official recall information

Official actions

According to the FDA recall records:

  • Customers were instructed to identify and quarantine affected kits and submit the recall response form, including affected inventory quantities.
  • Medline stated that over-labels would be supplied instructing staff to remove and discard the affected neurological sponge before using the kit.
  • Distributors and organizations that transferred affected kits were instructed to notify downstream customers and include their quantities on the response form.
Entry 07FDA Class II

AVID Medical recalls Halyard IV START kits containing affected BD ChloraPrep applicators

Halyard IV START KIT NON-STERILE models KRIV26-01 and KRIV26-02 containing BD ChloraPrep Clear 1 mL applicators

Status
Open, Classified
Recalling firm
AVID Medical, Inc.
Manufacturer
AVID Medical, Inc.
Event ID
99764
Recall number
Z-3260-2026, Z-3261-2026
FDA posting date
Recall initiated
Quantity
1,008 kits (21 cases) across two FDA recall records
Distribution
US Nationwide distribution.

Official recall information

What happened

FDA posted two Class II recall records for Halyard IV START kits containing BD ChloraPrep Clear 1 mL applicators that may have an open or incomplete packaging seal. An updated field correction also identified certain applicator lots contaminated with Aspergillus penicillioides. The two FDA records report compatible quantities totaling 1,008 kits, or 21 cases, distributed nationwide in the United States.

MedTech Current analysis

Why it matters

Facilities need to distinguish kits covered by the contamination update from kits affected only by the packaging-seal concern. The required response differs: some kits must be discarded, while others can be managed through quarantine and component removal.

Official recall information

Official actions

According to the FDA recall records:

  • For kits containing applicators affected by the contamination update, consignees were instructed to discard the impacted kits and notify customers that received them.
  • For kits included for sterility concerns but not the contamination update, consignees were instructed to quarantine affected product, discontinue use of the affected applicator component, and apply a warning label directing users to discard that component.
  • Customers were instructed to notify downstream recipients when affected kits had been further distributed.
Entry 08FDA Class II

Heraeus recalls PALACOS pro bone-cement systems over potential monomer leakage

PALACOS R pro 40/80, PALACOS R+G pro 40/80, PALACOS MV+G pro 40/80, and PALACOS MV pro 40/80 bone-cement systems

Status
Open, Classified
Recalling firm
Heraeus Medical GmbH (Dental Division)
Manufacturer
Heraeus Medical GmbH (Dental Division)
Event ID
99708
Recall number
Z-3301-2026, Z-3302-2026, Z-3303-2026, Z-3304-2026
FDA posting date
Recall initiated
Quantity
Z-3301-2026: 33,727 units; Z-3302-2026: 59,660 units; Z-3303-2026: 20,523 units; Z-3304-2026: quantity not reported
Distribution
Worldwide distribution, including U.S. nationwide distribution and listed international markets.

Official recall information

What happened

FDA posted four Class II recall records for PALACOS pro bone-cement systems because of potential monomer-liquid leakage. The affected families include standard and antibiotic-loaded PALACOS R and MV configurations. FDA reports 33,727 units, 59,660 units, and 20,523 units for three records; the fourth record does not report a Quantity in Commerce. Distribution is worldwide, including the United States.

MedTech Current analysis

Why it matters

Potential monomer leakage changes how staff handle these pre-filled cement systems. The notice calls for eye protection during operation and an immediate stop if leakage appears; the all-lot scope across four configurations also broadens inventory checks.

Official recall information

Official actions

According to the FDA recall records:

  • Consignees were instructed to ensure relevant hospital departments received the correction letter and to wear suitable eye protection whenever the system is operated.
  • If leakage is observed, users were instructed not to continue using the product and to discard it using additional precautions, including double gloves, according to hospital and local waste procedures.
  • Consignees were instructed to complete and return the reply form and retain a copy for their records.

Entries were selected by FDA Recall Posting Date from individual FDA recall database records; every retained FDA record URL is linked for traceability.