Device RecallsRecall Watch

Recall Watch · FDA Class II

FDA Class II Recall Watch: 14–19 September 2026

Eight selected FDA Class II recalls posted from 14–19 September 2026 highlight transcatheter-device integrity, anesthesia delivery interruption, orthopaedic implant material aging, imaging-data mismatches, diagnostic measurement errors, hemodialysis handling risks, and radiotherapy software limitations. The issue emphasizes both direct clinical consequences and the operational burden of software corrections, inventory control, product returns, and user notifications.

Published Issue FDA posting period 14–19 September 2026

At a glance

A compact index of this issue. Select an entry to move to its complete record.

Selected FDA Class II recall entries in this issue
Company/ProductRecall issueOfficial actionQuantityEntry link
PI CARDIA LTDPi-Cardia ShortCutPi-Cardia recalls ShortCut catheters over positioning-arm weld issuePi-Cardia stated that its team would contact affected customers with instructions to identify and return affected inventory.529 unitsPi-Cardia recalls ShortCut catheters over positioning-arm weld issue — read full entry
Datex-Ohmeda, Inc.Aisys CS2; Aisys CS2 with Et ControlDatex-Ohmeda recalls Aisys CS2 systems over alternate-oxygen transition issueUsers were instructed to complete the specified preoperative checkout and tests before use.147 units across two FDA recall recordsDatex-Ohmeda recalls Aisys CS2 systems over alternate-oxygen transition issue — read full entry
Howmedica Osteonics Corp.Stryker UHMWPE orthopaedic implant components used in hip, knee, shoulder, and elbow systemsStryker recalls UHMWPE orthopaedic implant components over material agingCustomers were instructed to identify affected product in inventory.12,156 US; same FDA Quantity in Commerce value repeated across 12 grouped recall recordsStryker recalls UHMWPE orthopaedic implant components over material aging — read full entry
GE HealthcareWorkflow Core Services; Core Services Hub V8GE Healthcare recalls imaging workflow systems over patient-data association riskCustomers were instructed not to begin new inbound DICOM migration projects without contacting GE Healthcare.419 systems across two FDA recall recordsGE Healthcare recalls imaging workflow systems over patient-data association risk — read full entry
Philips Ultrasound, LLCPhilips EPIQ Elite Diagnostic Ultrasound System with software version 14.0Philips recalls EPIQ Elite systems over Doppler sample-location errorUsers were instructed not to use the C5-1 Abd Vasc TSP when there is a possibility of a weak PW Doppler signal.219 unitsPhilips recalls EPIQ Elite systems over Doppler sample-location error — read full entry
Midmark CorporationIQvitals Zone configurations affected by Linear Deflation Method performance issueMidmark recalls IQvitals Zone configurations over inaccurate blood-pressure readingsUsers were instructed to use only the Step Deflation Method for blood-pressure measurement.207 units across four FDA recall recordsMidmark recalls IQvitals Zone configurations over inaccurate blood-pressure readings — read full entry
Fresenius Medical Care Renal Therapies Group, LLCCombiSet Hemodialysis Blood Tubing SetFresenius recalls CombiSet blood tubing over connection and handling riskCustomers were instructed to review the device instructions for use.1743026 unitsFresenius recalls CombiSet blood tubing over connection and handling risk — read full entry
PTW FREIBURGVERIQA RT MonteCarlo 3D v3, S070031.520, software version 3.1.18PTW recalls VERIQA over bolus-display limitation in secondary dose calculationUsers uncertain whether a bolus is required were instructed not to use VERIQA RT MonteCarlo 3D as the source for that determination.2PTW recalls VERIQA over bolus-display limitation in secondary dose calculation — read full entry

Entry 02

Datex-Ohmeda recalls Aisys CS2 systems over alternate-oxygen transition issue

Company/Product
Datex-Ohmeda, Inc. — Aisys CS2; Aisys CS2 with Et Control
Recall issue
Datex-Ohmeda recalls Aisys CS2 systems over alternate-oxygen transition issue
Official action
Users were instructed to complete the specified preoperative checkout and tests before use.
Quantity
147 units across two FDA recall records
Datex-Ohmeda recalls Aisys CS2 systems over alternate-oxygen transition issue — read full entry

Entry 03

Stryker recalls UHMWPE orthopaedic implant components over material aging

Company/Product
Howmedica Osteonics Corp. — Stryker UHMWPE orthopaedic implant components used in hip, knee, shoulder, and elbow systems
Recall issue
Stryker recalls UHMWPE orthopaedic implant components over material aging
Official action
Customers were instructed to identify affected product in inventory.
Quantity
12,156 US; same FDA Quantity in Commerce value repeated across 12 grouped recall records
Stryker recalls UHMWPE orthopaedic implant components over material aging — read full entry

Entry 04

GE Healthcare recalls imaging workflow systems over patient-data association risk

Company/Product
GE Healthcare — Workflow Core Services; Core Services Hub V8
Recall issue
GE Healthcare recalls imaging workflow systems over patient-data association risk
Official action
Customers were instructed not to begin new inbound DICOM migration projects without contacting GE Healthcare.
Quantity
419 systems across two FDA recall records
GE Healthcare recalls imaging workflow systems over patient-data association risk — read full entry

Entry 05

Philips recalls EPIQ Elite systems over Doppler sample-location error

Company/Product
Philips Ultrasound, LLC — Philips EPIQ Elite Diagnostic Ultrasound System with software version 14.0
Recall issue
Philips recalls EPIQ Elite systems over Doppler sample-location error
Official action
Users were instructed not to use the C5-1 Abd Vasc TSP when there is a possibility of a weak PW Doppler signal.
Quantity
219 units
Philips recalls EPIQ Elite systems over Doppler sample-location error — read full entry

Entry 06

Midmark recalls IQvitals Zone configurations over inaccurate blood-pressure readings

Company/Product
Midmark Corporation — IQvitals Zone configurations affected by Linear Deflation Method performance issue
Recall issue
Midmark recalls IQvitals Zone configurations over inaccurate blood-pressure readings
Official action
Users were instructed to use only the Step Deflation Method for blood-pressure measurement.
Quantity
207 units across four FDA recall records
Midmark recalls IQvitals Zone configurations over inaccurate blood-pressure readings — read full entry

Entry 08

PTW recalls VERIQA over bolus-display limitation in secondary dose calculation

Company/Product
PTW FREIBURG — VERIQA RT MonteCarlo 3D v3, S070031.520, software version 3.1.18
Recall issue
PTW recalls VERIQA over bolus-display limitation in secondary dose calculation
Official action
Users uncertain whether a bolus is required were instructed not to use VERIQA RT MonteCarlo 3D as the source for that determination.
Quantity
2
PTW recalls VERIQA over bolus-display limitation in secondary dose calculation — read full entry

This week’s entries

Official recall facts and actions are separated from MedTech Current significance analysis.

Entry 01FDA Class II

Pi-Cardia recalls ShortCut catheters over positioning-arm weld issue

Pi-Cardia ShortCut

Status
Open, Classified
Recalling firm
PI CARDIA LTD
Manufacturer
PI CARDIA LTD
Event ID
99664
Recall number
Z-3213-2026
FDA posting date
Recall initiated
Quantity
529 units
Distribution
U.S. distribution in UT, KS, CA, MA, PA, NH, DE, and NY.

Official recall information

What happened

FDA classified Pi-Cardia's ShortCut recall as Class II after the firm identified a weld issue that may allow the catheter's positioning arm to detach. ShortCut is used during transcatheter heart-valve procedures to facilitate leaflet modification. FDA lists 529 units in commerce across eight U.S. states. The firm instructed affected customers to identify and return impacted inventory.

MedTech Current analysis

Why it matters

A structural failure involving a catheter component used during a transcatheter valve procedure can create immediate procedural disruption and retrieval concerns. The event is also relevant to cardiovascular-device manufacturers because a localized weld-control problem can trigger inventory identification and return activity across multiple clinical sites.

Official recall information

Official actions

According to the FDA recall record:

  • Pi-Cardia stated that its team would contact affected customers with instructions to identify and return affected inventory.
  • Customers were directed to contact their Pi-Cardia representative with questions regarding the field action.
  • The firm stated that additional updates would be provided as necessary.
Entry 02FDA Class II

Datex-Ohmeda recalls Aisys CS2 systems over alternate-oxygen transition issue

Aisys CS2; Aisys CS2 with Et Control

Status
Open, Classified
Recalling firm
Datex-Ohmeda, Inc.
Manufacturer
Datex-Ohmeda, Inc.
Event ID
99649
Recall number
Z-3190-2026, Z-3191-2026
FDA posting date
Recall initiated
Quantity
147 units across two FDA recall records
Distribution
Worldwide - US Nationwide distribution.

Official recall information

What happened

FDA posted two Class II records for Aisys CS2 anesthesia systems after certain electronic gas-mixer assemblies were found capable of automatically transitioning to Alternate O2 control. The change delivers 100% oxygen and alarms, but anesthetic-agent delivery is interrupted until clinicians intervene. The grouped records cover Aisys CS2 and Aisys CS2 with Et Control, totaling 147 units in commerce.

MedTech Current analysis

Why it matters

An unexpected transition that interrupts anesthetic-agent delivery can demand immediate clinician recognition and response during anesthesia. The event is significant to hospitals because the correction combines interim operating instructions with inspection of affected service-stock components, while manufacturers must address a nonconforming component in an installed system population.

Official recall information

Official actions

According to the FDA recall records:

  • Users were instructed to complete the specified preoperative checkout and tests before use.
  • If Alternate O2 control occurs, clinicians were instructed to adjust oxygen flow and use the available guidance for anesthetic delivery.
  • If anesthetic-agent delivery is unavailable, users were instructed to supplement or transition to intravenous anesthetics as clinically appropriate.
  • Affected service-stock gas-mixer assemblies were to be inspected and disposed of, and GE stated that affected systems would be corrected at no cost.
Entry 03FDA Class II

Stryker recalls UHMWPE orthopaedic implant components over material aging

Stryker UHMWPE orthopaedic implant components used in hip, knee, shoulder, and elbow systems

Status
Open, Classified
Recalling firm
Howmedica Osteonics Corp.
Manufacturer
Howmedica Osteonics Corp.
Event ID
99558
Recall number
Z-3195-2026, Z-3196-2026, Z-3197-2026, Z-3198-2026, Z-3199-2026, Z-3200-2026, Z-3201-2026, Z-3202-2026, Z-3203-2026, Z-3204-2026, Z-3205-2026, Z-3206-2026
FDA posting date
Recall initiated
Quantity
12,156 US; same FDA Quantity in Commerce value repeated across 12 grouped recall records
Distribution
Worldwide distribution including US Nationwide, Canada, Latin America, Europe, Asia Pacific, and South Africa.

Official recall information

What happened

FDA posted 12 Class II records under event 99558 for Stryker orthopaedic implant components made with UHMWPE raw material aged more than five years. The firm identified potential elevated oxidation that may increase wear and contribute to wear debris, material fragments, or device fracture or dissociation. FDA repeats the same Quantity in Commerce value, 12,156 US, across the grouped records, so it is not summed.

MedTech Current analysis

Why it matters

The affected products are implanted orthopaedic components, and the FDA record links increased material oxidation to wear-related outcomes that may include pain, inflammation, and revision surgery. For manufacturers, the event also illustrates how raw-material aging controls and process-change governance can propagate across numerous implant configurations and product families.

Official recall information

Official actions

According to the FDA recall records:

  • Customers were instructed to identify affected product in inventory.
  • Affected product was to be quarantined and use discontinued.
  • Customers were instructed to return affected product to Stryker.
Entry 04FDA Class II

GE Healthcare recalls imaging workflow systems over patient-data association risk

Workflow Core Services; Core Services Hub V8

Status
Open, Classified
Recalling firm
GE Healthcare
Manufacturer
GE Healthcare
Event ID
99661
Recall number
Z-3170-2026, Z-3171-2026
FDA posting date
Recall initiated
Quantity
419 systems across two FDA recall records
Distribution
Worldwide - US Nationwide distribution.

Official recall information

What happened

FDA posted two Class II records for GE Healthcare Workflow Core Services and Core Services Hub V8 after identifying circumstances in which historical imaging data migrated from a legacy system may become associated with the wrong patient. Studies or images could then appear under an incorrect patient record, creating potential for misdiagnosis or delayed diagnosis. The grouped records total 419 systems worldwide.

MedTech Current analysis

Why it matters

Incorrect patient association is a data-integrity failure with direct diagnostic implications even when the underlying images themselves are unchanged. The recall is also operationally significant because healthcare organizations may need to preserve legacy systems, pause new DICOM migration work, and verify migrated studies against patient identifiers while awaiting correction.

Official recall information

Official actions

According to the FDA recall records:

  • Customers were instructed not to begin new inbound DICOM migration projects without contacting GE Healthcare.
  • Users were instructed to verify that studies and images are associated with the correct patient using available identifiers and clinical context.
  • Customers were instructed to maintain prior source systems or migration servers until GE provides further correction guidance.
  • Facilities were asked to circulate the notice and return the acknowledgement response.
Entry 05FDA Class II

Philips recalls EPIQ Elite systems over Doppler sample-location error

Philips EPIQ Elite Diagnostic Ultrasound System with software version 14.0

Status
Open, Classified
Recalling firm
Philips Ultrasound, LLC
Manufacturer
Philips Ultrasound, LLC
Event ID
99706
Recall number
Z-3249-2026
FDA posting date
Recall initiated
Quantity
219 units
Distribution
Worldwide distribution including US Nationwide.

Official recall information

What happened

FDA classified Philips EPIQ Elite systems running software version 14.0 as Class II because the displayed Doppler sample may appear shallower than the actual measurement location. The discrepancy can produce inaccurate blood-flow velocity measurements and may contribute to misdiagnosis. FDA lists 219 units in commerce, with distribution across the United States and multiple international markets.

MedTech Current analysis

Why it matters

Doppler measurements can influence vascular and hemodynamic assessment, so a mismatch between displayed and actual sample location has direct diagnostic relevance. The correction also illustrates how a software-design change can alter measurement interpretation while the device remains usable under specific interim restrictions communicated to clinical users.

Official recall information

Official actions

According to the FDA recall record:

  • Users were instructed not to use the C5-1 Abd Vasc TSP when there is a possibility of a weak PW Doppler signal.
  • Users were instructed to use another appropriate TSP for affected examinations.
  • Philips stated that other uses could continue according to the device instructions for use.
  • Facilities were asked to circulate and post the notice as appropriate and return the customer response form within 30 days.
Entry 06FDA Class II

Midmark recalls IQvitals Zone configurations over inaccurate blood-pressure readings

IQvitals Zone configurations affected by Linear Deflation Method performance issue

Status
Open, Classified
Recalling firm
Midmark Corporation
Manufacturer
Midmark Corporation
Event ID
99641
Recall number
Z-3217-2026, Z-3218-2026, Z-3219-2026, Z-3220-2026
FDA posting date
Recall initiated
Quantity
207 units across four FDA recall records
Distribution
US Nationwide distribution.

Official recall information

What happened

FDA posted four Class II records for Midmark IQvitals Zone configurations because a performance issue may produce inaccurate blood-pressure readings when the Linear Deflation Method is used. The records share the same event and corrective action, so they are grouped here as one editorial event. FDA lists 1, 153, 20, and 33 units respectively, totaling 207 units in commerce.

MedTech Current analysis

Why it matters

Inaccurate blood-pressure measurements can affect clinical assessment and treatment decisions, especially when the error is not obvious to the user. The field action is operationally important because facilities must verify the measurement method on affected devices and temporarily standardize use around the Step Deflation setting until the product correction is available.

Official recall information

Official actions

According to the FDA recall records:

  • Users were instructed to use only the Step Deflation Method for blood-pressure measurement.
  • Users were instructed not to use the Linear Deflation Method and to verify that Step Deflation is selected.
  • Facilities were asked to inform affected users and complete and return the response form.
  • Midmark stated that a product correction would be provided.
Entry 07FDA Class II

Fresenius recalls CombiSet blood tubing over connection and handling risk

CombiSet Hemodialysis Blood Tubing Set

Status
Open, Classified
Recalling firm
Fresenius Medical Care Renal Therapies Group, LLC
Manufacturer
Fresenius Medical Care Renal Therapies Group, LLC
Event ID
99700
Recall number
Z-3225-2026
FDA posting date
Recall initiated
Quantity
1743026 units
Distribution
Domestic: Nationwide Distribution.

Official recall information

What happened

FDA classified the CombiSet Hemodialysis Blood Tubing Set action as Class II and states that the communication recommends renewed attention to the instructions for use and treatment-preparation steps. The record identifies potential consequences including blood loss, treatment delay, and air entering the system if connections are not properly secured. FDA lists 1743026 units in commerce with nationwide U.S. distribution.

MedTech Current analysis

Why it matters

Hemodialysis blood tubing is part of an extracorporeal blood circuit, so connection integrity and preparation steps have direct clinical and workflow importance. The event is notable because the field action focuses on reinforcing correct use rather than removing product, placing much of the immediate burden on training, setup verification, and clinical practice.

Official recall information

Official actions

According to the FDA recall record:

  • Customers were instructed to review the device instructions for use.
  • Users were instructed to ensure that all treatment-preparation and handling steps are performed as directed.
  • The notice emphasizes ensuring that transducer-protector and female-luer connections are finger tight to maintain a closed system.
Entry 08FDA Class II

PTW recalls VERIQA over bolus-display limitation in secondary dose calculation

VERIQA RT MonteCarlo 3D v3, S070031.520, software version 3.1.18

Status
Open, Classified
Recalling firm
PTW FREIBURG
Manufacturer
PTW FREIBURG
Event ID
99720
Recall number
Z-3174-2026
FDA posting date
Recall initiated
Quantity
2
Distribution
Worldwide distribution including the U.S. states of GA and NY and multiple international markets.

Official recall information

What happened

FDA classified PTW's VERIQA RT MonteCarlo 3D v3 recall as Class II because bolus material included in a treatment plan is considered in secondary dose calculation but is not clearly displayed in the software. A user could therefore omit the bolus during treatment and deliver dose to the wrong tissue depth. FDA reports a Quantity in Commerce of 2 without specifying a unit.

MedTech Current analysis

Why it matters

The issue affects radiotherapy dose-checking software and can create a mismatch between calculated and delivered treatment setup if the bolus requirement is missed. Although the quantity is small, the consequence is clinically meaningful because the limitation concerns delivered dose depth.

Official recall information

Official actions

According to the FDA recall record:

  • Users uncertain whether a bolus is required were instructed not to use VERIQA RT MonteCarlo 3D as the source for that determination.
  • Users were instructed to rely on the treatment-planning system and patient record for bolus requirements.
  • The firm stated that VERIQA should not influence the clinical workflow or treatment plan for this determination.
  • PTW stated that the issue would be corrected in a future software release and asked recipients to forward the notice as appropriate.

Entries were selected by FDA Recall Posting Date from individual FDA recall database records; each retained FDA record URL is preserved for traceability.