Device RecallsRecall Watch

Recall Watch · FDA Class II

FDA Class II Recall Watch: 30 August–5 September 2026

Eight selected FDA Class II recall events for 30 August–5 September 2026 highlight hemodialysis blood-leak and dialysate-contamination risks, infusion-pump software, robotic-surgery motion, delayed cardiac-monitor alerts, unintended CT radiation, contaminated skin-prep components, and ophthalmic tissue-injury concerns. The actions span stop-use, quarantine, software mitigation, field correction, component removal, return, replacement, and user notification.

Published Issue FDA posting period 30 August–5 September 2026

At a glance

A compact index of this issue. Select an entry to move to its complete record.

Selected FDA Class II recall entries in this issue
Company/ProductRecall issueOfficial actionQuantityEntry link
Fresenius Medical Care Renal Therapies Group, LLC5008X Standard HD Pre-Flush Standard Blood Tubing Set, Part Number 03-5110-6Fresenius recalls specified 5008X blood tubing sets over blood-leak riskCustomers were instructed to identify affected product on hand and immediately segregate and discontinue use of affected lots.Standard HD Pre-Flush: 120 units; HD/HDF Standard: 62,544 units; HD/HDF with CLiC: 18,600; HD/HDF with Twister: 6,624 unitsFresenius recalls specified 5008X blood tubing sets over blood-leak risk — read full entry
Baxter Healthcare CorporationBaxter Novum IQ Syringe Pump, Model 40800BAXUSBaxter recalls Novum IQ syringe pumps over multi-syringe infusion software issueCustomers were advised to use the same compatible syringe model, brand, and size for all syringes required within a multi-syringe Volume/Time workflow.14,605 units across two FDA recall recordsBaxter recalls Novum IQ syringe pumps over multi-syringe infusion software issue — read full entry
Medos International SarlVELYS Enabling Tech Robotic-Assist Station, REF 451611001VELYS Robotic-Assist Station recalled over potential unintended robotic-arm movementConsignees were instructed not to use the VELYS Robotic-Assist Station and to quarantine it immediately.29 stationsVELYS Robotic-Assist Station recalled over potential unintended robotic-arm movement — read full entry
Boston Scientific CorporationLUX-Dx, LUX-Dx II, and LUX-Dx II+ Insertable Cardiac Monitor System server software for U.S., EU, and ANZ LATITUDE Clarity deploymentsBoston Scientific recalls LUX-Dx server software over delayed AF Burden alertsUntil updated server software is available, customers may mitigate the delay by changing the AF Burden Alert threshold or enabling AF Events alerts with a minimum duration.Not reportedBoston Scientific recalls LUX-Dx server software over delayed AF Burden alerts — read full entry
Fresenius Medical Care Renal Therapies Group, LLCGranuFlo Dry Acid ConcentrateFresenius recalls specified GranuFlo, NaturaLyte, and Citrasate lots over discolorationCustomers were instructed to identify affected lots on hand and immediately segregate and discontinue use of those lots.GranuFlo: 27,216 cs; 2,316 ea; NaturaLyte: 2,998 cs; 68 ea; Citrasate: 56,036 cs; 1,788 drums; 204 eaFresenius recalls specified GranuFlo, NaturaLyte, and Citrasate lots over discoloration — read full entry
Philips North America LlcPhilips Brilliance iCTPhilips recalls four CT product families over unintended radiation issuePhilips is informing customers through an Electronic Product Radiation Correction notification.Not reportedPhilips recalls four CT product families over unintended radiation issue — read full entry
Medline Industries, LPMedline Convenience Kits containing specified BD ChloraPrep Clear 1mL Applicators across multiple procedure-kit categoriesMedline recalls convenience kits containing affected BD ChloraPrep applicatorsMedline instructed customers to request stickers to over-label affected convenience kits.222,653 totalMedline recalls convenience kits containing affected BD ChloraPrep applicators — read full entry
Johnson & Johnson Surgical Vision, Inc.HEALON EndoCoat PRO Ophthalmic Viscosurgical DeviceJohnson & Johnson Surgical Vision corrects HEALON products over insertion-related tissue-injury riskUsers were instructed to withdraw the cannula and reassess insertion conditions if resistance is encountered rather than forcing insertion.111,478Johnson & Johnson Surgical Vision corrects HEALON products over insertion-related tissue-injury risk — read full entry

Entry 01

Fresenius recalls specified 5008X blood tubing sets over blood-leak risk

Company/Product
Fresenius Medical Care Renal Therapies Group, LLC — 5008X Standard HD Pre-Flush Standard Blood Tubing Set, Part Number 03-5110-6
Recall issue
Fresenius recalls specified 5008X blood tubing sets over blood-leak risk
Official action
Customers were instructed to identify affected product on hand and immediately segregate and discontinue use of affected lots.
Quantity
Standard HD Pre-Flush: 120 units; HD/HDF Standard: 62,544 units; HD/HDF with CLiC: 18,600; HD/HDF with Twister: 6,624 units
Fresenius recalls specified 5008X blood tubing sets over blood-leak risk — read full entry

Entry 02

Baxter recalls Novum IQ syringe pumps over multi-syringe infusion software issue

Company/Product
Baxter Healthcare Corporation — Baxter Novum IQ Syringe Pump, Model 40800BAXUS
Recall issue
Baxter recalls Novum IQ syringe pumps over multi-syringe infusion software issue
Official action
Customers were advised to use the same compatible syringe model, brand, and size for all syringes required within a multi-syringe Volume/Time workflow.
Quantity
14,605 units across two FDA recall records
Baxter recalls Novum IQ syringe pumps over multi-syringe infusion software issue — read full entry

Entry 03

VELYS Robotic-Assist Station recalled over potential unintended robotic-arm movement

Company/Product
Medos International Sarl — VELYS Enabling Tech Robotic-Assist Station, REF 451611001
Recall issue
VELYS Robotic-Assist Station recalled over potential unintended robotic-arm movement
Official action
Consignees were instructed not to use the VELYS Robotic-Assist Station and to quarantine it immediately.
Quantity
29 stations
VELYS Robotic-Assist Station recalled over potential unintended robotic-arm movement — read full entry

Entry 04

Boston Scientific recalls LUX-Dx server software over delayed AF Burden alerts

Company/Product
Boston Scientific Corporation — LUX-Dx, LUX-Dx II, and LUX-Dx II+ Insertable Cardiac Monitor System server software for U.S., EU, and ANZ LATITUDE Clarity deployments
Recall issue
Boston Scientific recalls LUX-Dx server software over delayed AF Burden alerts
Official action
Until updated server software is available, customers may mitigate the delay by changing the AF Burden Alert threshold or enabling AF Events alerts with a minimum duration.
Quantity
Not reported
Boston Scientific recalls LUX-Dx server software over delayed AF Burden alerts — read full entry

Entry 05

Fresenius recalls specified GranuFlo, NaturaLyte, and Citrasate lots over discoloration

Company/Product
Fresenius Medical Care Renal Therapies Group, LLC — GranuFlo Dry Acid Concentrate
Recall issue
Fresenius recalls specified GranuFlo, NaturaLyte, and Citrasate lots over discoloration
Official action
Customers were instructed to identify affected lots on hand and immediately segregate and discontinue use of those lots.
Quantity
GranuFlo: 27,216 cs; 2,316 ea; NaturaLyte: 2,998 cs; 68 ea; Citrasate: 56,036 cs; 1,788 drums; 204 ea
Fresenius recalls specified GranuFlo, NaturaLyte, and Citrasate lots over discoloration — read full entry

Entry 07

Medline recalls convenience kits containing affected BD ChloraPrep applicators

Company/Product
Medline Industries, LP — Medline Convenience Kits containing specified BD ChloraPrep Clear 1mL Applicators across multiple procedure-kit categories
Recall issue
Medline recalls convenience kits containing affected BD ChloraPrep applicators
Official action
Medline instructed customers to request stickers to over-label affected convenience kits.
Quantity
222,653 total
Medline recalls convenience kits containing affected BD ChloraPrep applicators — read full entry

Entry 08

Johnson & Johnson Surgical Vision corrects HEALON products over insertion-related tissue-injury risk

Company/Product
Johnson & Johnson Surgical Vision, Inc. — HEALON EndoCoat PRO Ophthalmic Viscosurgical Device
Recall issue
Johnson & Johnson Surgical Vision corrects HEALON products over insertion-related tissue-injury risk
Official action
Users were instructed to withdraw the cannula and reassess insertion conditions if resistance is encountered rather than forcing insertion.
Quantity
111,478
Johnson & Johnson Surgical Vision corrects HEALON products over insertion-related tissue-injury risk — read full entry

This week’s entries

Official recall facts and actions are separated from MedTech Current significance analysis.

Entry 01FDA Class II

Fresenius recalls specified 5008X blood tubing sets over blood-leak risk

5008X Standard HD Pre-Flush Standard Blood Tubing Set, Part Number 03-5110-6; 5008X HD/HDF Standard Blood Tubing Set, Part Number 03-5300-3; 5008X HD/HDF with CLiC Blood Tubing Set, Part Number 03-5300-3C; 5008X HD/HDF with Twister Blood Tubing Set, Part Number 03-5350-8

Status
Open, Classified
Recalling firm
Fresenius Medical Care Renal Therapies Group, LLC
Manufacturer
Fresenius Medical Care Renal Therapies Group, LLC
Event ID
99556
Recall number
Z-3084-2026, Z-3085-2026, Z-3086-2026, Z-3087-2026
FDA posting date
Recall initiated
Quantity
Standard HD Pre-Flush: 120 units; HD/HDF Standard: 62,544 units; HD/HDF with CLiC: 18,600; HD/HDF with Twister: 6,624 units
Distribution
Worldwide distribution, including U.S. nationwide distribution and Lithuania.

Official recall information

What happened

FDA posted four Class II recall records for Fresenius 5008X blood tubing sets after identifying a heightened probability of blood leakage at the connection between the Arterial Alpha Clip and the main line or pump segment. A leak during hemodialysis may cause blood loss and hypovolemia. Because one FDA quantity lacks a stated unit, the four official quantities are preserved separately rather than aggregated.

MedTech Current analysis

Why it matters

Blood-circuit leakage during hemodialysis has direct treatment and patient-safety implications and can require immediate inventory segregation across dialysis programs. The event also demonstrates why quantity aggregation must respect FDA's stated units: otherwise a seemingly simple total can introduce unsupported assumptions into an otherwise evidence-based recall summary.

Official recall information

Official actions

According to the FDA recall records:

  • Customers were instructed to identify affected product on hand and immediately segregate and discontinue use of affected lots.
  • Customers were instructed to contact the designated call center for return of affected lots and replacement stock.
  • Customers were instructed to complete and return the included reply form.
Entry 02FDA Class II

Baxter recalls Novum IQ syringe pumps over multi-syringe infusion software issue

Baxter Novum IQ Syringe Pump, Model 40800BAXUS; Baxter Novum IQ Syringe Pump, Loaner, Model 40800BAXUSL

Status
Open, Classified
Recalling firm
Baxter Healthcare Corporation
Manufacturer
Baxter Healthcare Corporation
Event ID
99557
Recall number
Z-3065-2026, Z-3066-2026
FDA posting date
Recall initiated
Quantity
14,605 units across two FDA recall records
Distribution
U.S. nationwide distribution.

Official recall information

What happened

FDA posted two Class II recall records for Baxter Novum IQ Syringe Pumps covering standard and loaner models. The software issue can occur during a Volume/Time multi-syringe infusion when an empty syringe is replaced while a Volume To Be Infused remains. FDA reports 14,601 standard pumps and 4 loaner pumps, totaling 14,605 units across the two records.

MedTech Current analysis

Why it matters

Infusion-pump software behavior during syringe changes can affect medication-delivery workflow at the bedside, making the event particularly relevant to clinical engineering and medication-safety teams. The correction also depends on users following a specific compatible-syringe workflow until the underlying software behavior is addressed.

Official recall information

Official actions

According to the FDA recall records:

  • Customers were advised to use the same compatible syringe model, brand, and size for all syringes required within a multi-syringe Volume/Time workflow.
  • Customers were provided additional instructions for situations in which a syringe does not support the intended flow rate after a syringe change.
Entry 03FDA Class II

VELYS Robotic-Assist Station recalled over potential unintended robotic-arm movement

VELYS Enabling Tech Robotic-Assist Station, REF 451611001

Status
Open, Classified
Recalling firm
Medos International Sarl
Event ID
99467
Recall number
Z-3088-2026
FDA posting date
Recall initiated
Quantity
29 stations
Distribution
U.S. nationwide distribution in multiple states.

Official recall information

What happened

FDA classified the recall of the VELYS Enabling Tech Robotic-Assist Station as Class II after a software issue in the robotic controller was identified that may cause unintended robotic-arm movement during bone preparation or instrument positioning. The affected scope is 29 stations distributed across multiple U.S. states. The associated navigation and camera stations may continue to be used under the notice.

MedTech Current analysis

Why it matters

Unintended motion during orthopedic robotic procedures has obvious procedural implications and creates a high-priority field response despite the relatively small installed base. The event also distinguishes the recalled robotic-assist component from related navigation and camera equipment that the official notice permits facilities to continue using.

Official recall information

Official actions

According to the FDA recall record:

  • Consignees were instructed not to use the VELYS Robotic-Assist Station and to quarantine it immediately.
  • The VELYS Navigation Station and VELYS Camera Station may continue to be used according to their instructions for use.
  • Consignees were instructed to complete and return the business response form and forward or post the notice for relevant personnel.
Entry 04FDA Class II

Boston Scientific recalls LUX-Dx server software over delayed AF Burden alerts

LUX-Dx, LUX-Dx II, and LUX-Dx II+ Insertable Cardiac Monitor System server software for U.S., EU, and ANZ LATITUDE Clarity deployments

Status
Open, Classified
Recalling firm
Boston Scientific Corporation
Manufacturer
Boston Scientific Corporation
Event ID
99524
Recall number
Z-3081-2026, Z-3082-2026, Z-3083-2026
FDA posting date
Recall initiated
Quantity
Not reported
Distribution
Worldwide distribution, including U.S. nationwide distribution.

Official recall information

What happened

FDA posted three Class II records for LATITUDE Clarity server software supporting the LUX-Dx family of insertable cardiac monitors. The software can delay AF Burden Alert notification when a non-default threshold is configured above zero hours per day, which may delay clinical evaluation or therapeutic intervention for an atrial arrhythmia. The implanted monitors continue detecting, storing, and transmitting data.

MedTech Current analysis

Why it matters

This event separates device sensing from downstream alert delivery: the implant may function as intended while server-side software delays the notification clinicians rely on. That makes the recall a useful example of how remote-monitoring infrastructure can create clinically meaningful risk without requiring replacement of the implanted device itself.

Official recall information

Official actions

According to the FDA recall records:

  • Until updated server software is available, customers may mitigate the delay by changing the AF Burden Alert threshold or enabling AF Events alerts with a minimum duration.
  • Boston Scientific does not recommend replacement of the implanted insertable cardiac monitor.
  • Customers were asked to share the notification with healthcare professionals managing patients through the LUX-Dx and LATITUDE Clarity system.
Entry 05FDA Class II

Fresenius recalls specified GranuFlo, NaturaLyte, and Citrasate lots over discoloration

GranuFlo Dry Acid Concentrate; NaturaLyte Liquid Acid Concentrate; Citrasate Liquid Acid Concentrate

Status
Open, Classified
Recalling firm
Fresenius Medical Care Renal Therapies Group, LLC
Manufacturer
Fresenius Medical Care Renal Therapies Group, LLC
Event ID
99543
Recall number
Z-3091-2026, Z-3092-2026, Z-3093-2026
FDA posting date
Recall initiated
Quantity
GranuFlo: 27,216 cs; 2,316 ea; NaturaLyte: 2,998 cs; 68 ea; Citrasate: 56,036 cs; 1,788 drums; 204 ea
Distribution
Worldwide distribution, including U.S. nationwide distribution and Mexico.

Official recall information

What happened

FDA posted three Class II records for specified GranuFlo, NaturaLyte, and Citrasate hemodialysis concentrate lots following complaints of blue-green product discoloration. FDA identified the cause as material or component contamination. Because the records use multiple quantity units, including cases, eaches, and drums, the official quantities are preserved by product rather than combined into a single total.

MedTech Current analysis

Why it matters

Dialysate concentrate is a core treatment input, so visible product contamination or discoloration requires rapid inventory control in dialysis operations. The event is also a practical example of why grouped recall entries should preserve incompatible units instead of converting or summing them in ways the FDA record does not support.

Official recall information

Official actions

According to the FDA recall records:

  • Customers were instructed to identify affected lots on hand and immediately segregate and discontinue use of those lots.
  • Customers were instructed to contact Fresenius for return of affected lots and replacement stock.
  • Customers were instructed to complete and return the attached reply form.
Entry 06FDA Class II

Philips recalls four CT product families over unintended radiation issue

Philips Brilliance iCT; Philips IQon Spectral CT; Philips Ingenuity CT; Philips Brilliance CT 64

Status
Open, Classified
Recalling firm
Philips North America Llc
Manufacturer
Philips North America Llc
Event ID
99784
Recall number
Z-3040-2026, Z-3041-2026, Z-3042-2026, Z-3043-2026
FDA posting date
Recall initiated
Quantity
Not reported
Distribution
United States and U.S. territories.

Official recall information

What happened

FDA posted four Class II recall records covering Philips Brilliance iCT, IQon Spectral CT, Ingenuity CT, and Brilliance CT 64 systems. Philips identified an unintended radiation issue that could affect equipment performance. The records share the same FDA Event ID, recall reason, customer correction, and posting date, so the four closely related CT product records are grouped into one editorial entry.

MedTech Current analysis

Why it matters

A radiation-related field correction across several CT platforms has broad relevance for imaging departments, service organizations, and device quality teams. The shared software-correction approach also makes grouping appropriate: readers can understand one underlying field action without losing the individual model families, recall numbers, or source records.

Official recall information

Official actions

According to the FDA recall records:

  • Philips is informing customers through an Electronic Product Radiation Correction notification.
  • Philips will schedule appointments with remaining customers to update affected system software free of charge.
  • Customers needing additional information or support were directed to their local Philips representative or customer-care center.
Entry 07FDA Class II

Medline recalls convenience kits containing affected BD ChloraPrep applicators

Medline Convenience Kits containing specified BD ChloraPrep Clear 1mL Applicators across multiple procedure-kit categories

Status
Open, Classified
Recalling firm
Medline Industries, LP
Manufacturer
Medline Industries, LP
Event ID
99561
Recall number
Z-3067-2026, Z-3068-2026, Z-3069-2026, Z-3070-2026, Z-3071-2026, Z-3072-2026, Z-3073-2026, Z-3074-2026, Z-3075-2026, Z-3076-2026
FDA posting date
Recall initiated
Quantity
222,653 total
Distribution
U.S. nationwide distribution.

Official recall information

What happened

FDA posted ten Class II records for Medline convenience kits containing specified BD ChloraPrep Clear 1mL Applicators. The component supplier recalled affected applicators because packaging may have an open or incomplete seal, creating contamination and infection concerns. Each FDA record repeats the same Quantity in Commerce value of 222,653 total, so that event-level figure is preserved once rather than multiplied across records.

MedTech Current analysis

Why it matters

A supplier packaging defect can propagate through many downstream procedure-kit configurations, creating substantial traceability and relabeling work for hospitals and distributors. The event also illustrates an important quantity-handling edge case: an identical event total repeated across multiple FDA records should not be summed as though each were a separate quantity.

Official recall information

Official actions

According to the FDA recall records:

  • Medline instructed customers to request stickers to over-label affected convenience kits.
  • The over-label instructs users to remove the affected ChloraPrep component and replace it with product from supply.
  • The FDA records state that the affected convenience kits are not to be returned.
Entry 08FDA Class II

Johnson & Johnson Surgical Vision corrects HEALON products over insertion-related tissue-injury risk

HEALON EndoCoat PRO Ophthalmic Viscosurgical Device; HEALON DUET PRO II Dual Pack

Status
Open, Classified
Recalling firm
Johnson & Johnson Surgical Vision, Inc.
Manufacturer
Johnson & Johnson Surgical Vision, Inc.
Event ID
99496
Recall number
Z-3064-2026
FDA posting date
Recall initiated
Quantity
111,478
Distribution
Worldwide distribution, including U.S. nationwide distribution.

Official recall information

What happened

FDA classified a correction involving HEALON EndoCoat PRO and HEALON DUET PRO II as Class II. The notice clarifies insertion technique because forcing the cannula when resistance is encountered may increase the risk of tissue injury, including corneal abrasion, iris trauma, or Descemet's membrane detachment. FDA lists 111,478 affected products in commerce across U.S. and international distribution.

MedTech Current analysis

Why it matters

This is a labeling and technique-focused correction rather than a product-removal event, but the potential consequences involve direct ophthalmic tissue injury during cataract procedures. It is a useful reminder that field corrections may address device-design interaction with user technique even when products remain available for use under clarified instructions.

Official recall information

Official actions

According to the FDA recall record:

  • Users were instructed to withdraw the cannula and reassess insertion conditions if resistance is encountered rather than forcing insertion.
  • Customers were instructed to forward the notice to surgeons using the device and to other facilities or surgeons that received affected product.
  • Customers were instructed to keep a copy of the notice with affected product and return the business reply form.

Entries were selected by FDA Recall Posting Date from individual FDA recall database records; the supplied source records provide the underlying event and action details.