FDA Class II Recall Watch: 23–29 August 2026

Eight selected FDA Class II recalls for 23–29 August 2026 highlight ECG waveform errors, patient-monitoring software failures, hemodialysis alarms, operating-table lock failures, loose interventional X-ray hardware, inaccurate PACS measurements, endotoxin-related surgical-kit actions, and false cleaning-indicator results. The issue emphasizes software correction, device inspection, quarantine, and field-service workload.

Published Issue FDA posting period 23–29 August 2026

At a glance

A compact index of this issue. Select an entry to move to its complete record.

Selected FDA Class II recall entries in this issue
Company/ProductRecall issueOfficial actionQuantityEntry link
Midmark CorporationMidmark IQecg Digital ECG, models 4-000-0061 and 4-000-0062Midmark recalls Digital ECG systems over waveform and heart-rate software anomalyUntil corrected software is installed, Midmark recommends a manual clinical assessment of heart rate when ECG tracings reflect tachycardia before therapeutic intervention.3,624 units reported as the combined quantity for both FDA recall recordsMidmark recalls Digital ECG systems over waveform and heart-rate software anomaly — read full entry
GE Medical Systems Information Technologies IncPortrait Core Services Software V1.1GE Healthcare recalls Portrait software versions over monitoring loss and alarm resetsAffected Portrait systems may continue to be used.23 units; 4 units; 15 units; 0 across four FDA recall recordsGE Healthcare recalls Portrait software versions over monitoring loss and alarm resets — read full entry
Fresenius Medical Care Renal Therapies Group, LLCFresenius Medical Care 2008T BlueStar Hemodialysis Machine, models 191126 and 191130Fresenius recalls specified 2008T BlueStar hemodialysis machines following alarm complaintsAffected machines in use without error messages may continue to be used until the parts are replaced.1140 systemsFresenius recalls specified 2008T BlueStar hemodialysis machines following alarm complaints — read full entry
IMRIS Imaging IncORT200 operating room tableIMRIS recalls ORT200 and ORT300 tables over rotational-lock mechanism failureFacilities were instructed to test affected tables daily in standard positions and inspect for anomalies.19 units across two FDA recall recordsIMRIS recalls ORT200 and ORT300 tables over rotational-lock mechanism failure — read full entry
Canon Medical System, USA, INC.Alphenix interventional fluoroscopic X-ray systems, models INFX-8000C, INFX-8000F and INFX-8000VCanon recalls specified Alphenix systems over loose ceiling-arm fixing screwsIf abnormal noise occurs or a sensor error is displayed before correction, customers were instructed to stop using the device and contact Canon service.10 systemsCanon recalls specified Alphenix systems over loose ceiling-arm fixing screws — read full entry
FUJIFILM Healthcare Americas CorporationFUJIFILM Synapse PACS software v7.4.310FUJIFILM recalls Synapse PACS v7.4.310 over inaccurate ruler measurementsCustomers were instructed to post the notification on or near the Synapse PACS workstation.6 unitsFUJIFILM recalls Synapse PACS v7.4.310 over inaccurate ruler measurements — read full entry
AVID Medical, Inc.Halyard medical convenience kits across general surgery, neurological and orthopedic configurations containing affected Ritmed neurological spongesAvid Medical recalls Halyard kits containing neurological sponges with endotoxin variabilityConsignees were instructed to identify, segregate, and quarantine impacted product.28,802 kits across three FDA recall recordsAvid Medical recalls Halyard kits containing neurological sponges with endotoxin variability — read full entry
Healthmark Industries Co., Inc.healthmark ProChek-II, models PT-313 and PT-201-12Healthmark recalls ProChek protein indicators manufactured with expired ingredientFacilities were instructed to ensure affected product is no longer in use and is properly quarantined.5,143 units across two FDA recall recordsHealthmark recalls ProChek protein indicators manufactured with expired ingredient — read full entry

Entry 01

Midmark recalls Digital ECG systems over waveform and heart-rate software anomaly

Company/Product
Midmark Corporation — Midmark IQecg Digital ECG, models 4-000-0061 and 4-000-0062
Recall issue
Midmark recalls Digital ECG systems over waveform and heart-rate software anomaly
Official action
Until corrected software is installed, Midmark recommends a manual clinical assessment of heart rate when ECG tracings reflect tachycardia before therapeutic intervention.
Quantity
3,624 units reported as the combined quantity for both FDA recall records
Midmark recalls Digital ECG systems over waveform and heart-rate software anomaly — read full entry

Entry 02

GE Healthcare recalls Portrait software versions over monitoring loss and alarm resets

Company/Product
GE Medical Systems Information Technologies Inc — Portrait Core Services Software V1.1
Recall issue
GE Healthcare recalls Portrait software versions over monitoring loss and alarm resets
Official action
Affected Portrait systems may continue to be used.
Quantity
23 units; 4 units; 15 units; 0 across four FDA recall records
GE Healthcare recalls Portrait software versions over monitoring loss and alarm resets — read full entry

Entry 03

Fresenius recalls specified 2008T BlueStar hemodialysis machines following alarm complaints

Company/Product
Fresenius Medical Care Renal Therapies Group, LLC — Fresenius Medical Care 2008T BlueStar Hemodialysis Machine, models 191126 and 191130
Recall issue
Fresenius recalls specified 2008T BlueStar hemodialysis machines following alarm complaints
Official action
Affected machines in use without error messages may continue to be used until the parts are replaced.
Quantity
1140 systems
Fresenius recalls specified 2008T BlueStar hemodialysis machines following alarm complaints — read full entry

Entry 04

IMRIS recalls ORT200 and ORT300 tables over rotational-lock mechanism failure

Company/Product
IMRIS Imaging Inc — ORT200 operating room table
Recall issue
IMRIS recalls ORT200 and ORT300 tables over rotational-lock mechanism failure
Official action
Facilities were instructed to test affected tables daily in standard positions and inspect for anomalies.
Quantity
19 units across two FDA recall records
IMRIS recalls ORT200 and ORT300 tables over rotational-lock mechanism failure — read full entry

Entry 05

Canon recalls specified Alphenix systems over loose ceiling-arm fixing screws

Company/Product
Canon Medical System, USA, INC. — Alphenix interventional fluoroscopic X-ray systems, models INFX-8000C, INFX-8000F and INFX-8000V
Recall issue
Canon recalls specified Alphenix systems over loose ceiling-arm fixing screws
Official action
If abnormal noise occurs or a sensor error is displayed before correction, customers were instructed to stop using the device and contact Canon service.
Quantity
10 systems
Canon recalls specified Alphenix systems over loose ceiling-arm fixing screws — read full entry

Entry 07

Avid Medical recalls Halyard kits containing neurological sponges with endotoxin variability

Company/Product
AVID Medical, Inc. — Halyard medical convenience kits across general surgery, neurological and orthopedic configurations containing affected Ritmed neurological sponges
Recall issue
Avid Medical recalls Halyard kits containing neurological sponges with endotoxin variability
Official action
Consignees were instructed to identify, segregate, and quarantine impacted product.
Quantity
28,802 kits across three FDA recall records
Avid Medical recalls Halyard kits containing neurological sponges with endotoxin variability — read full entry

Entry 08

Healthmark recalls ProChek protein indicators manufactured with expired ingredient

Company/Product
Healthmark Industries Co., Inc. — healthmark ProChek-II, models PT-313 and PT-201-12
Recall issue
Healthmark recalls ProChek protein indicators manufactured with expired ingredient
Official action
Facilities were instructed to ensure affected product is no longer in use and is properly quarantined.
Quantity
5,143 units across two FDA recall records
Healthmark recalls ProChek protein indicators manufactured with expired ingredient — read full entry

This week’s entries

Official recall facts and actions are separated from MedTech Current significance analysis.

Entry 01FDA Class II

Midmark recalls Digital ECG systems over waveform and heart-rate software anomaly

Midmark IQecg Digital ECG, models 4-000-0061 and 4-000-0062; Midmark Digital ECG HD+12 and HD+15, model 4-000-0080

Status
Open, Classified
Recalling firm
Midmark Corporation
Manufacturer
Midmark Corporation
Event ID
99404
Recall number
Z-3012-2026, Z-3013-2026
FDA posting date
Recall initiated
Quantity
3,624 units reported as the combined quantity for both FDA recall records
Distribution
Worldwide distribution - US Nationwide and Canada.

Official recall information

What happened

FDA posted two Class II records under event 99404 for Midmark IQecg and Digital ECG systems using IQecg Plugin versions starting with v10 and before v10.1.0.18. A communication anomaly may produce incorrect electrocardiogram waveforms and an inaccurate beats-per-minute calculation. FDA states that the reported 3,624-unit quantity is combined across the IQecg and Midmark Digital ECG systems.

MedTech Current analysis

Why it matters

Incorrect ECG waveforms or heart-rate values can affect clinical interpretation and therapeutic decisions, making this a high-relevance software correction. The event also shows why manufacturers need clear interim verification steps when a software defect can alter a displayed physiological signal before a corrected version is installed.

Official recall information

Official actions

According to the FDA recall records:

  • Until corrected software is installed, Midmark recommends a manual clinical assessment of heart rate when ECG tracings reflect tachycardia before therapeutic intervention.
  • Users encountering the issue were directed to follow the temporary workaround instructions provided in the notification.
  • Midmark will contact facilities to verify installed IQecg Plugin versions and coordinate the software correction.
Entry 02FDA Class II

GE Healthcare recalls Portrait software versions over monitoring loss and alarm resets

Portrait Core Services Software V1.1; Portrait Compute and Core Services V1.2; Portrait Core Services Software V1.0.5; FRU Portrait Core Services SW Reload V1.1

Status
Open, Classified
Recalling firm
GE Medical Systems Information Technologies Inc
Manufacturer
GE Medical Systems Information Technologies Inc
Event ID
99536
Recall number
Z-3036-2026, Z-3037-2026, Z-3038-2026, Z-3039-2026
FDA posting date
Recall initiated
Quantity
23 units; 4 units; 15 units; 0 across four FDA recall records
Distribution
United States distribution in Michigan and New York, with distribution to multiple international markets.

Official recall information

What happened

FDA posted four Class II records under event 99536 for GE Healthcare Portrait Core Services software configurations. The software issue may cause loss of patient monitoring at the Hub and Central Viewer, and alarm settings may reset to factory defaults after an update rather than retain configured patient or care-area settings. FDA reports quantities of 23 units, 4 units, 15 units, and 0 across the four records.

MedTech Current analysis

Why it matters

Loss of centralized monitoring and reset alarm settings can create immediate clinical and operational consequences, especially during a software update. The event is also a strong configuration-management example: a correction can introduce risk when patient-specific or care-area alarm settings do not persist as users expect.

Official recall information

Official actions

According to the FDA recall records:

  • Affected Portrait systems may continue to be used.
  • Facilities were instructed to establish appropriate alternative monitoring before a software update is started.
  • Facilities were instructed to place the supplied User Manual addendum with the Portrait system documentation.
  • Potential users were to be made aware of the safety notification and recommended actions.
Entry 03FDA Class II

Fresenius recalls specified 2008T BlueStar hemodialysis machines following alarm complaints

Fresenius Medical Care 2008T BlueStar Hemodialysis Machine, models 191126 and 191130

Status
Open, Classified
Recalling firm
Fresenius Medical Care Renal Therapies Group, LLC
Manufacturer
Fresenius Medical Care Renal Therapies Group, LLC
Event ID
99438
Recall number
Z-3017-2026
FDA posting date
Recall initiated
Quantity
1140 systems
Distribution
US Nationwide distribution.

Official recall information

What happened

FDA classified the correction of specified Fresenius 2008T BlueStar Hemodialysis Machines as Class II after the firm received complaints involving flow-, no-water-, or temperature-related alarms. The affected products are models 191126 and 191130. FDA lists 1140 systems in commerce with nationwide U.S. distribution, and Fresenius is arranging replacement of affected parts through its technical-service organization.

MedTech Current analysis

Why it matters

Hemodialysis machines support ongoing therapy, so recurring flow, water, or temperature alarms can impose immediate operational burden on clinics even when a machine remains usable without error messages. The correction also requires facilities to coordinate backup-machine use and field replacement according to each affected system's condition.

Official recall information

Official actions

According to the FDA recall record:

  • Affected machines in use without error messages may continue to be used until the parts are replaced.
  • If an affected machine develops flow-, no-water-, or temperature-related alarms, facilities were instructed to use a backup machine until parts are replaced.
  • Parts on affected machines not yet installed at clinics are to be replaced before installation.
  • Facilities were instructed to contact Fresenius technical service to schedule the update.
Entry 04FDA Class II

IMRIS recalls ORT200 and ORT300 tables over rotational-lock mechanism failure

ORT200 operating room table; ORT300 operating room table

Status
Open, Classified
Recalling firm
IMRIS Imaging Inc
Manufacturer
IMRIS Imaging Inc
Event ID
99596
Recall number
Z-3033-2026, Z-3034-2026
FDA posting date
Recall initiated
Quantity
19 units across two FDA recall records
Distribution
Worldwide distribution - US Nationwide and Canada, China, Japan, and Sweden.

Official recall information

What happened

FDA posted two Class II records under event 99596 for IMRIS ORT200 and ORT300 operating tables. The linear actuator and rotational-lock mechanism may mechanically fail, preventing transition between locked and unlocked states and creating a risk of unintended rotational movement during a procedure. FDA lists 6 ORT200 units and 13 ORT300 units, totaling 19 units across the two records.

MedTech Current analysis

Why it matters

Unintended movement of an operating table during a procedure creates a direct procedural safety concern even though the affected installed base is small. The recall is also operationally significant because facilities must perform interim daily checks and locked-position verification until field service replaces the affected components with an updated design.

Official recall information

Official actions

According to the FDA recall records:

  • Facilities were instructed to test affected tables daily in standard positions and inspect for anomalies.
  • During use, staff were instructed to verify that the table is locked by manually checking for rotational movement.
  • If the floor-latching mechanism does not function properly, facilities were instructed to contact IMRIS Customer Service.
  • IMRIS field service will replace the linear-actuator and rotational-lock components with an updated design.
Entry 05FDA Class II

Canon recalls specified Alphenix systems over loose ceiling-arm fixing screws

Alphenix interventional fluoroscopic X-ray systems, models INFX-8000C, INFX-8000F and INFX-8000V

Status
Open, Classified
Recalling firm
Canon Medical System, USA, INC.
Manufacturer
Canon Medical System, USA, INC.
Event ID
99222
Recall number
Z-3014-2026
FDA posting date
Recall initiated
Quantity
10 systems
Distribution
U.S. Nationwide distribution in CA, GA, MN, NC, NV, PA, OH, and TX.

Official recall information

What happened

FDA classified Canon Medical's correction of specified Alphenix interventional fluoroscopic X-ray systems as Class II because fixing screws in the ceiling-movement gear of the ceiling-suspended arm may become loose. Abnormal noise or sensor errors can indicate loosening. FDA lists 10 affected systems distributed in eight U.S. states, and Canon plans to inspect screw tightening and attachment through field service.

MedTech Current analysis

Why it matters

A mechanical issue in a ceiling-suspended interventional imaging system can interrupt procedures and requires prompt recognition of warning signs. Although only 10 systems are listed, the event is professionally relevant because it combines procedural imaging, mechanical hardware, stop-use criteria, and a service-based inspection and correction program.

Official recall information

Official actions

According to the FDA recall record:

  • If abnormal noise occurs or a sensor error is displayed before correction, customers were instructed to stop using the device and contact Canon service.
  • Facilities were strongly requested to share the notice with users and clinical engineering or biomedical personnel.
  • Canon representatives will contact customers to schedule inspection of screw tightening and proper attachment.
Entry 06FDA Class II

FUJIFILM recalls Synapse PACS v7.4.310 over inaccurate ruler measurements

FUJIFILM Synapse PACS software v7.4.310

Status
Open, Classified
Recalling firm
FUJIFILM Healthcare Americas Corporation
Manufacturer
FUJIFILM Healthcare Americas Corporation
Event ID
99529
Recall number
Z-3030-2026
FDA posting date
Recall initiated
Quantity
6 units
Distribution
Worldwide - US distribution in NY, VA, KS, OR, and OH, and distribution in Australia.

Official recall information

What happened

FDA classified FUJIFILM Synapse PACS software v7.4.310 as a Class II recall because a software issue may make ruler measurements inaccurate. The affected radiological image-processing system has six units in commerce, distributed in five U.S. states and Australia. FUJIFILM provided awareness instructions and a workaround allowing M-mode ruler measurements to be performed on the ultrasound system before study transfer.

MedTech Current analysis

Why it matters

Measurement accuracy is central to diagnostic imaging workflows, so even a small installed base can justify attention when software may alter a displayed measurement. The correction also illustrates a common software-field-action pattern: users may continue operating the system when they understand the limitation and apply an approved workaround.

Official recall information

Official actions

According to the FDA recall record:

  • Customers were instructed to post the notification on or near the Synapse PACS workstation.
  • Customers were instructed to complete and return the Field Action Verification Form.
  • Customers were directed to contact the Technical Assistance Center to open a service ticket.
  • The notice describes performing M-mode ruler measurements on the ultrasound system and transferring the study to Synapse PACS as a workaround.
Entry 07FDA Class II

Avid Medical recalls Halyard kits containing neurological sponges with endotoxin variability

Halyard medical convenience kits across general surgery, neurological and orthopedic configurations containing affected Ritmed neurological sponges

Status
Open, Classified
Recalling firm
AVID Medical, Inc.
Manufacturer
AVID Medical, Inc.
Event ID
99469
Recall number
Z-3044-2026, Z-3045-2026, Z-3046-2026
FDA posting date
Recall initiated
Quantity
28,802 kits across three FDA recall records
Distribution
US Nationwide distribution across multiple states.

Official recall information

What happened

FDA posted three Class II records under event 99469 for Halyard medical convenience kits manufactured by Avid Medical. The kits contain specified Ritmed non-sterile neurological sponges recalled because of potential unexpected variability in endotoxin levels. FDA reports 11,763 general-surgery kits, 16,223 neurological kits, and 816 orthopedic kits, totaling 28,802 kits across the three records.

MedTech Current analysis

Why it matters

A component-level endotoxin issue can propagate across many finished surgical-kit configurations and create substantial inventory and labeling work for healthcare organizations. The event is a useful supply-chain example because the field action focuses on identifying affected kits, controlling them in inventory, and removing the recalled component rather than discarding every complete kit.

Official recall information

Official actions

According to the FDA recall records:

  • Consignees were instructed to identify, segregate, and quarantine impacted product.
  • Recall labels were to be added to affected Avid kits so users would discard the impacted component.
  • Distributors that further distributed affected kits were instructed to notify end users and maintain effectiveness records.
Entry 08FDA Class II

Healthmark recalls ProChek protein indicators manufactured with expired ingredient

healthmark ProChek-II, models PT-313 and PT-201-12; healthmark ProChek-W Protein Test for Robotic Arms and ProChek-Semi Quantitative Protein Test, model PT-SQ-001

Status
Open, Classified
Recalling firm
Healthmark Industries Co., Inc.
Manufacturer
Healthmark Industries Co., Inc.
Event ID
99537
Recall number
Z-3031-2026, Z-3032-2026
FDA posting date
Recall initiated
Quantity
5,143 units across two FDA recall records
Distribution
Worldwide distribution - US Nationwide and Canada, France, Hong Kong, and South Korea.

Official recall information

What happened

FDA posted two Class II records under event 99537 for Healthmark ProChek protein cleaning-process indicators. The tests were manufactured with an expired ingredient that may present as either a false-negative or false-positive result. FDA reports 5,090 units for ProChek-II and 53 units for ProChek-W/ProChek-Semi Quantitative, totaling 5,143 units across the two recall records.

MedTech Current analysis

Why it matters

Protein indicators are used to assess cleaning processes, so false-negative or false-positive results can undermine confidence in reprocessing verification and create unnecessary follow-up work. The event is particularly relevant to quality teams because an expired manufacturing input translated directly into performance uncertainty in a process-monitoring product.

Official recall information

Official actions

According to the FDA recall records:

  • Facilities were instructed to ensure affected product is no longer in use and is properly quarantined.
  • The notice was to be forwarded to relevant individuals and organizations that received transferred product.
  • Customers were instructed to contact Healthmark Customer Service to return unused affected product.
  • Customers were instructed to complete the Customer Response Form.