FDA Class II Recall Watch: 28 September–4 October 2026

Eight selected FDA Class II recalls posted from 28 September through 4 October 2026 highlight intracranial implant mislabeling, insulin-pump software failure, cardiopulmonary bypass leakage, CT imaging and radiation issues, false laboratory results, sterilization-monitoring failures, and surgical-positioner weld defects. The actions span quarantine, product return, software correction, laboratory review, destruction, replacement, and field inspection.

Published Issue FDA posting period 28 September–4 October 2026

At a glance

A compact index of this issue. Select an entry to move to its complete record.

Selected FDA Class II recall entries in this issue
Company/ProductRecall issueOfficial actionQuantityEntry link
Micro Therapeutics, Inc.Pipeline Flex Embolization Device with Shield Technology, REF PED2-300-12, lot D112860Micro Therapeutics recalls Pipeline Flex lot over incorrect implant diameter labelingCustomers were instructed not to use affected devices and to remove and quarantine unused affected product.2Micro Therapeutics recalls Pipeline Flex lot over incorrect implant diameter labeling — read full entry
Tandem Diabetes Care, Inc.Tandem t:slim mobile application, iOS software version 2.13 (TestFlight)Tandem recalls beta t:slim mobile app after Libre 3 Plus pairing failureCustomers were advised to discontinue use of iOS t:slim mobile application version 2.13.177Tandem recalls beta t:slim mobile app after Libre 3 Plus pairing failure — read full entry
SPECTRUM MEDICAL SRLQuantum PureFlow Standard Heat Exchanger High Flow 3/8, model HX55W-S0Spectrum Medical recalls Quantum PureFlow heat exchanger over blood-to-glycol leakageUsers were instructed to monitor the heat exchanger during procedures for a color change in the glycol compartment indicating leakage.3232Spectrum Medical recalls Quantum PureFlow heat exchanger over blood-to-glycol leakage — read full entry
Philips North America LlcSpectral CT 7500, software version 5.2.x, product code 728333Philips recalls Spectral CT 7500 over Bolus Tracking issue and potential additional radiationPhilips issued an Electronic Product Radiation Correction notification describing actions intended to prevent patient risk.62Philips recalls Spectral CT 7500 over Bolus Tracking issue and potential additional radiation — read full entry
Siemens Healthcare Diagnostics, Inc.Atellica CH Revised C-Reactive Protein (RCRP), Siemens Material Number 11537223, all lotsSiemens recalls Atellica CH RCRP assay over potential falsely elevated resultsCustomers were instructed to follow the supplied instructions to set up and calibrate the device.42940 units (3938 US, 39002 OUS)Siemens recalls Atellica CH RCRP assay over potential falsely elevated results — read full entry
Steris CorporationSTERIS VERIFY biological indicators and test packsSTERIS recalls VERIFY biological indicators and test packs over low spore populationCustomers were instructed to destroy any affected product remaining on hand.LCB007: 409; S3060: 1185; S3061: 2205; S3065: 217; S3069: 233STERIS recalls VERIFY biological indicators and test packs over low spore population — read full entry
PHILIPS MEDICAL SYSTEMSPhilips Spectral CT and Spectral CT 7500 RTPhilips recalls Spectral CT systems over image, slice-thickness, and motion software issuesCustomers were told they may continue using affected systems in accordance with their intended use.539Philips recalls Spectral CT systems over image, slice-thickness, and motion software issues — read full entry
QUANTUM OPS IncADAPTABLE Beach Chair surgical patient positioner, product numbers 804-33-300 and 3105-400-030Quantum Ops recalls ADAPTABLE Beach Chair over head-rest weld failureCustomers were instructed to locate affected devices and perform the specified safety check.266Quantum Ops recalls ADAPTABLE Beach Chair over head-rest weld failure — read full entry

Entry 01

Micro Therapeutics recalls Pipeline Flex lot over incorrect implant diameter labeling

Company/Product
Micro Therapeutics, Inc. — Pipeline Flex Embolization Device with Shield Technology, REF PED2-300-12, lot D112860
Recall issue
Micro Therapeutics recalls Pipeline Flex lot over incorrect implant diameter labeling
Official action
Customers were instructed not to use affected devices and to remove and quarantine unused affected product.
Quantity
2
Micro Therapeutics recalls Pipeline Flex lot over incorrect implant diameter labeling — read full entry

Entry 02

Tandem recalls beta t:slim mobile app after Libre 3 Plus pairing failure

Company/Product
Tandem Diabetes Care, Inc. — Tandem t:slim mobile application, iOS software version 2.13 (TestFlight)
Recall issue
Tandem recalls beta t:slim mobile app after Libre 3 Plus pairing failure
Official action
Customers were advised to discontinue use of iOS t:slim mobile application version 2.13.
Quantity
177
Tandem recalls beta t:slim mobile app after Libre 3 Plus pairing failure — read full entry

Entry 03

Spectrum Medical recalls Quantum PureFlow heat exchanger over blood-to-glycol leakage

Company/Product
SPECTRUM MEDICAL SRL — Quantum PureFlow Standard Heat Exchanger High Flow 3/8, model HX55W-S0
Recall issue
Spectrum Medical recalls Quantum PureFlow heat exchanger over blood-to-glycol leakage
Official action
Users were instructed to monitor the heat exchanger during procedures for a color change in the glycol compartment indicating leakage.
Quantity
3232
Spectrum Medical recalls Quantum PureFlow heat exchanger over blood-to-glycol leakage — read full entry

Entry 04

Philips recalls Spectral CT 7500 over Bolus Tracking issue and potential additional radiation

Company/Product
Philips North America Llc — Spectral CT 7500, software version 5.2.x, product code 728333
Recall issue
Philips recalls Spectral CT 7500 over Bolus Tracking issue and potential additional radiation
Official action
Philips issued an Electronic Product Radiation Correction notification describing actions intended to prevent patient risk.
Quantity
62
Philips recalls Spectral CT 7500 over Bolus Tracking issue and potential additional radiation — read full entry

Entry 05

Siemens recalls Atellica CH RCRP assay over potential falsely elevated results

Company/Product
Siemens Healthcare Diagnostics, Inc. — Atellica CH Revised C-Reactive Protein (RCRP), Siemens Material Number 11537223, all lots
Recall issue
Siemens recalls Atellica CH RCRP assay over potential falsely elevated results
Official action
Customers were instructed to follow the supplied instructions to set up and calibrate the device.
Quantity
42940 units (3938 US, 39002 OUS)
Siemens recalls Atellica CH RCRP assay over potential falsely elevated results — read full entry

Entry 06

STERIS recalls VERIFY biological indicators and test packs over low spore population

Company/Product
Steris Corporation — STERIS VERIFY biological indicators and test packs
Recall issue
STERIS recalls VERIFY biological indicators and test packs over low spore population
Official action
Customers were instructed to destroy any affected product remaining on hand.
Quantity
LCB007: 409; S3060: 1185; S3061: 2205; S3065: 217; S3069: 233
STERIS recalls VERIFY biological indicators and test packs over low spore population — read full entry

Entry 07

Philips recalls Spectral CT systems over image, slice-thickness, and motion software issues

Company/Product
PHILIPS MEDICAL SYSTEMS — Philips Spectral CT and Spectral CT 7500 RT
Recall issue
Philips recalls Spectral CT systems over image, slice-thickness, and motion software issues
Official action
Customers were told they may continue using affected systems in accordance with their intended use.
Quantity
539
Philips recalls Spectral CT systems over image, slice-thickness, and motion software issues — read full entry

Entry 08

Quantum Ops recalls ADAPTABLE Beach Chair over head-rest weld failure

Company/Product
QUANTUM OPS Inc — ADAPTABLE Beach Chair surgical patient positioner, product numbers 804-33-300 and 3105-400-030
Recall issue
Quantum Ops recalls ADAPTABLE Beach Chair over head-rest weld failure
Official action
Customers were instructed to locate affected devices and perform the specified safety check.
Quantity
266
Quantum Ops recalls ADAPTABLE Beach Chair over head-rest weld failure — read full entry

This week’s entries

Official recall facts and actions are separated from MedTech Current significance analysis.

Entry 01FDA Class II

Micro Therapeutics recalls Pipeline Flex lot over incorrect implant diameter labeling

Pipeline Flex Embolization Device with Shield Technology, REF PED2-300-12, lot D112860

Status
Open, Classified
Recalling firm
Micro Therapeutics, Inc.
Event ID
99689
Recall number
Z-3326-2026
FDA posting date
Recall initiated
Quantity
2
Distribution
U.S. Nationwide distribution in the states of GA and TX.
Identifiers
REF PED2-300-12; Lot D112860

Official recall information

What happened

FDA classified the recall of Pipeline Flex Embolization Device with Shield Technology lot D112860 as Class II because devices labeled as 3.00 mm diameter contain 2.50 mm diameter implants. The product is a permanent intracranial aneurysm flow-diverter implant with a guidewire-based delivery system. FDA lists a Quantity in Commerce of 2, without a stated unit, distributed in Georgia and Texas.

MedTech Current analysis

Why it matters

The labeled diameter guides implant selection in a neurovascular procedure. A small reported quantity does not remove the concern: the implant remains permanently in the patient. The response therefore covers unused stock and patients already treated, whose records need review to assess whether follow-up is appropriate.

Official recall information

Official actions

According to the FDA recall record:

  • Customers were instructed not to use affected devices and to remove and quarantine unused affected product.
  • Affected product was to be returned to Medtronic, with replacement support available through the Medtronic representative.
  • Customers were instructed to review records for patients treated with the affected lot and assess whether follow-up is appropriate.
  • Customers were instructed to complete the confirmation form and share the recall communication with other affected organizations.
Entry 02FDA Class II

Tandem recalls beta t:slim mobile app after Libre 3 Plus pairing failure

Tandem t:slim mobile application, iOS software version 2.13 (TestFlight)

Status
Open, Classified
Recalling firm
Tandem Diabetes Care, Inc.
Event ID
99589
Recall number
Z-3330-2026
FDA posting date
Recall initiated
Quantity
177
Distribution
US Nationwide distribution.
Identifiers
iOS version 2.13 (TestFlight)

Official recall information

What happened

FDA classified Tandem's beta t:slim mobile application version 2.13 for iOS as Class II after a software issue prevented pairing with Abbott FreeStyle Libre 3 Plus sensors. Affected users may lose CGM values in the app and automated closed-loop operation, requiring manual pump use and potentially contributing to hypoglycemia or hyperglycemia. FDA lists a Quantity in Commerce of 177, without a stated unit.

MedTech Current analysis

Why it matters

Losing CGM connectivity changes how the insulin pump operates: users must manage delivery in manual mode without the app’s glucose values or automated control. For connected-device teams, the failure links release controls and sensor-compatibility testing to the ability to restore a working production version quickly.

Official recall information

Official actions

According to the FDA recall record:

  • Customers were advised to discontinue use of iOS t:slim mobile application version 2.13.
  • Customers were instructed to uninstall version 2.13 from their mobile device.
  • Customers were instructed to return to production version 2.12.1 from the Apple App Store.
  • Users were instructed to scan a compatible FreeStyle Libre 3 Plus sensor with version 2.12.1 when starting the next sensor session.
Entry 03FDA Class II

Spectrum Medical recalls Quantum PureFlow heat exchanger over blood-to-glycol leakage

Quantum PureFlow Standard Heat Exchanger High Flow 3/8, model HX55W-S0, lot 18051160301178

Status
Open, Classified
Recalling firm
SPECTRUM MEDICAL SRL
Event ID
99462
Recall number
Z-3328-2026
FDA posting date
Recall initiated
Quantity
3232
Distribution
US Nationwide distribution in the states of AL, CA, FL, KS, MA, MD, MI, MO, NC, NJ, NY, OH, OR, PA, SC, TN, TX, VA, and WI.
Identifiers
Model HX55W-S0; Lot 18051160301178

Official recall information

What happened

FDA classified the Quantum PureFlow Standard Heat Exchanger High Flow 3/8 recall as Class II following blood-to-glycol leakage between compartments during cardiopulmonary bypass. The single-use device heats or cools circulating blood through a compatible heater-cooler. The recall covers model HX55W-S0, lot 18051160301178. FDA lists 3232, without a quantity unit, and identifies process design as the cause.

MedTech Current analysis

Why it matters

A leak during bypass requires a response in the circuit and to the equipment supporting it. Replacing the disposable heat exchanger does not resolve contamination of the heater-cooler; that equipment also needs removal from clinical use, decontamination, and inspection.

Official recall information

Official actions

According to the FDA recall record:

  • Users were instructed to monitor the heat exchanger during procedures for a color change in the glycol compartment indicating leakage.
  • If leakage is observed, the heat exchanger should be immediately replaced in the circuit and returned to Spectrum Medical.
  • Remove a contaminated heater-cooler from clinical use and return it for decontamination and inspection; do not use it with another patient.
  • Customers using a third-party heater-cooler were instructed to remove it from clinical use and contact its manufacturer if blood contamination occurs.
Entry 04FDA Class II

Philips recalls Spectral CT 7500 over Bolus Tracking issue and potential additional radiation

Spectral CT 7500, software version 5.2.x, product code 728333

Status
Open, Classified
Recalling firm
Philips North America Llc
Event ID
99865
Recall number
Z-3155-2026
FDA posting date
Recall initiated
Quantity
62
Distribution
U.S. and U.S. Territories.
Identifiers
Product code 728333; Software version 5.2.x

Official recall information

What happened

FDA classified a Philips Spectral CT 7500 correction as Class II because the Bolus Tracking scan application may crash and Bolus Tracker shots may continue after the contrast threshold is reached. FDA states that this can result in additional radiation exposure to the patient. The affected configuration uses software version 5.2.x, and FDA lists a Quantity in Commerce of 62, without a stated unit.

MedTech Current analysis

Why it matters

The software failure can change radiation delivered during a scan, beyond disrupting system availability. Imaging departments must manage the affected workflow while the correction is arranged. For software-quality teams, it shows how failure to stop a scan sequence can become a radiation-control problem.

Official recall information

Official actions

According to the FDA recall record:

  • Philips issued an Electronic Product Radiation Correction notification describing actions intended to prevent patient risk.
  • Philips stated that it will schedule appointments with affected customers to install a software solution free of charge.
Entry 05FDA Class II

Siemens recalls Atellica CH RCRP assay over potential falsely elevated results

Atellica CH Revised C-Reactive Protein (RCRP), Siemens Material Number 11537223, all lots

Status
Open, Classified
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Event ID
99726
Recall number
Z-3335-2026
FDA posting date
Recall initiated
Quantity
42940 units (3938 US, 39002 OUS)
Distribution
Worldwide distribution, including U.S. nationwide distribution and multiple international markets.
Identifiers
Siemens Material Number 11537223; All lots

Official recall information

What happened

FDA classified Siemens' Atellica CH Revised C-Reactive Protein assay recall as Class II because of the potential for falsely elevated results. The recall applies to all lots of Siemens Material Number 11537223. FDA reports 42940 units in commerce, including 3938 in the United States and 39002 outside the United States, with worldwide distribution and the cause still under investigation.

MedTech Current analysis

Why it matters

An operational analyzer can produce results that mislead clinical interpretation. Beyond calibration and discarded tests, the Medical Director must decide whether earlier results need action. Laboratory and IVD quality teams face an assay correction and a review of its clinical implications.

Official recall information

Official actions

According to the FDA recall record:

  • Customers were instructed to follow the supplied instructions to set up and calibrate the device.
  • Customers were instructed to record unused tests discarded for reimbursement or credit.
  • Laboratories were instructed to review the letter with their Medical Director and determine the appropriate course of action, including for previously generated results if applicable.
  • Return the Field Correction Effectiveness Check Form within 30 days, retain the letter with laboratory records, and forward it to recipients of affected product.
Entry 06FDA Class II

STERIS recalls VERIFY biological indicators and test packs over low spore population

VERIFY LCB007, S3065, and S3069: lots 260810, 260810A, 261106, 270116; VERIFY S3060 and S3061: those lots plus 270302. Model-specific details and expiration dates appear in the five linked FDA records.

Status
Open, Classified
Recalling firm
Steris Corporation
Event ID
99704
Recall number
Z-0071-2027, Z-0072-2027, Z-0073-2027, Z-0074-2027, Z-0075-2027
FDA posting date
Recall initiated
Quantity
LCB007: 409; S3060: 1185; S3061: 2205; S3065: 217; S3069: 233
Distribution
Worldwide - US Nationwide distribution.
Identifiers
LCB007, S3065, S3069: lots 260810, 260810A, 261106, 270116; S3060, S3061: lots 260810, 260810A, 261106, 270116, 270302

Official recall information

What happened

FDA posted five Class II records under event 99704 for STERIS VERIFY biological indicators, test packs, and challenge packs. The Geobacillus stearothermophilus spore population may fall below the minimum shelf-life specification, potentially producing a no-growth result even when the steam cycle did not achieve the intended spore reduction. FDA reports record-level quantities of 409, 1185, 2205, 217, and 233 without stated units.

MedTech Current analysis

Why it matters

A no-growth result could give sterile-processing staff false reassurance that a steam cycle was effective. Affected stock must be identified and destroyed, while replacement depends on remaining shelf life. The five records belong together because they cover the same product family, failure, firm, and corrective instructions.

Official recall information

Official actions

According to the FDA recall records:

  • Customers were instructed to destroy any affected product remaining on hand.
  • STERIS stated that it will replace affected product that is still within its shelf life.
Entry 07FDA Class II

Philips recalls Spectral CT systems over image, slice-thickness, and motion software issues

Specified serial numbers in the FDA record: model 728333 in the U.S. and outside the U.S.; models 728340, 728344, and 728345 outside the U.S. Software 5.2.x in the U.S.; specified 5.0.x, 5.2.x, and 5.4.x versions outside the U.S.

Status
Open, Classified
Recalling firm
PHILIPS MEDICAL SYSTEMS
Event ID
99796
Recall number
Z-0078-2027
FDA posting date
Recall initiated
Quantity
539
Distribution
Worldwide - US Nationwide distribution and multiple international markets.
Identifiers
Models 728333, 728340, 728344, 728345; Affected serial numbers: see FDA Code Information

Official recall information

What happened

FDA posted a separate Class II Philips Spectral CT correction on October 2. Unlike the September 29 radiation-control recall, this event concerns software issues that may produce image artifacts, wrong slice thickness, or unintended system motion across Spectral CT and Spectral CT 7500 RT configurations. FDA lists a Quantity in Commerce of 539, without a stated unit, with worldwide distribution.

MedTech Current analysis

Why it matters

Image artifacts and wrong slice thickness can affect interpretation; unintended motion also concerns system behavior. Radiology, clinical engineering, and software change-control teams need the instructions for this event. It remains separate from the Bolus Tracking recall because its failure modes and corrective instructions differ.

Official recall information

Official actions

According to the FDA recall record:

  • Customers were told they may continue using affected systems in accordance with their intended use.
  • Facilities were instructed to circulate the notice to all users so they are aware of the issues, potential hazards, and manufacturer recommendations.
  • Facilities were instructed to post the notice near affected units until a solution is installed.
  • Customers were instructed to complete and return the response form within 30 days.
Entry 08FDA Class II

Quantum Ops recalls ADAPTABLE Beach Chair over head-rest weld failure

ADAPTABLE Beach Chair surgical patient positioner, product numbers 804-33-300 and 3105-400-030; only affected lots and serial numbers listed in FDA Code Information, with the stated demo/undelivered exclusions

Status
Open, Classified
Recalling firm
QUANTUM OPS Inc
Event ID
99384
Recall number
Z-0076-2027
FDA posting date
Recall initiated
Quantity
266
Distribution
U.S. Nationwide distribution in the states of CA and TX.
Identifiers
Product numbers 804-33-300 and 3105-400-030; UDI-DI 00190446667278; Affected lots and serial numbers: see FDA Code Information

Official recall information

What happened

FDA classified the ADAPTABLE Beach Chair surgical patient positioner recall as Class II because an inadequate weld in the patient head-positioning system can fracture below the head rest. The firm received complaints describing welded-joint fracture and head-rest separation. FDA lists a Quantity in Commerce of 266, without a stated unit, distributed in California and Texas, and identifies device design as the determined cause.

MedTech Current analysis

Why it matters

The positioner supports the patient’s upper body and head during surgery; a failed joint can disrupt positioning. Facilities must locate and check affected devices. Continued use depends on passing the safety check. Passing devices still need exchange for an upgraded unit.

Official recall information

Official actions

According to the FDA recall record:

  • Customers were instructed to locate affected devices and perform the specified safety check.
  • Devices that pass the safety check may remain in clinical use pending exchange for an upgraded unit.
  • Devices that fail the safety check must be segregated and removed from use, with return and replacement arranged.
  • Notify staff and organizations that received the product. Maintain awareness until actions are complete; return the acknowledgement or reply form even if no affected product is found.

This issue retains the supplied 28 September–4 October 2026 window and includes records by FDA Recall Posting Date. Eight selected events are represented by 12 linked FDA records; the five STERIS records are grouped by firm, product family, failure, and corrective action. Quantities without stated units are preserved without added units or aggregation.