Recall Watch · FDA Class II
FDA Class II Recall Watch: 28 September–4 October 2026
Eight selected FDA Class II recalls posted from 28 September through 4 October 2026 highlight intracranial implant mislabeling, insulin-pump software failure, cardiopulmonary bypass leakage, CT imaging and radiation issues, false laboratory results, sterilization-monitoring failures, and surgical-positioner weld defects. The actions span quarantine, product return, software correction, laboratory review, destruction, replacement, and field inspection.
Published Issue FDA posting period 28 September–4 October 2026
Weekly overview
At a glance
A compact index of this issue. Select an entry to move to its complete record.
| Company/Product | Recall issue | Official action | Quantity | Entry link |
|---|---|---|---|---|
| Micro Therapeutics, Inc.Pipeline Flex Embolization Device with Shield Technology, REF PED2-300-12, lot D112860 | Micro Therapeutics recalls Pipeline Flex lot over incorrect implant diameter labeling | Customers were instructed not to use affected devices and to remove and quarantine unused affected product. | 2 | Micro Therapeutics recalls Pipeline Flex lot over incorrect implant diameter labeling — read full entry |
| Tandem Diabetes Care, Inc.Tandem t:slim mobile application, iOS software version 2.13 (TestFlight) | Tandem recalls beta t:slim mobile app after Libre 3 Plus pairing failure | Customers were advised to discontinue use of iOS t:slim mobile application version 2.13. | 177 | Tandem recalls beta t:slim mobile app after Libre 3 Plus pairing failure — read full entry |
| SPECTRUM MEDICAL SRLQuantum PureFlow Standard Heat Exchanger High Flow 3/8, model HX55W-S0 | Spectrum Medical recalls Quantum PureFlow heat exchanger over blood-to-glycol leakage | Users were instructed to monitor the heat exchanger during procedures for a color change in the glycol compartment indicating leakage. | 3232 | Spectrum Medical recalls Quantum PureFlow heat exchanger over blood-to-glycol leakage — read full entry |
| Philips North America LlcSpectral CT 7500, software version 5.2.x, product code 728333 | Philips recalls Spectral CT 7500 over Bolus Tracking issue and potential additional radiation | Philips issued an Electronic Product Radiation Correction notification describing actions intended to prevent patient risk. | 62 | Philips recalls Spectral CT 7500 over Bolus Tracking issue and potential additional radiation — read full entry |
| Siemens Healthcare Diagnostics, Inc.Atellica CH Revised C-Reactive Protein (RCRP), Siemens Material Number 11537223, all lots | Siemens recalls Atellica CH RCRP assay over potential falsely elevated results | Customers were instructed to follow the supplied instructions to set up and calibrate the device. | 42940 units (3938 US, 39002 OUS) | Siemens recalls Atellica CH RCRP assay over potential falsely elevated results — read full entry |
| Steris CorporationSTERIS VERIFY biological indicators and test packs | STERIS recalls VERIFY biological indicators and test packs over low spore population | Customers were instructed to destroy any affected product remaining on hand. | LCB007: 409; S3060: 1185; S3061: 2205; S3065: 217; S3069: 233 | STERIS recalls VERIFY biological indicators and test packs over low spore population — read full entry |
| PHILIPS MEDICAL SYSTEMSPhilips Spectral CT and Spectral CT 7500 RT | Philips recalls Spectral CT systems over image, slice-thickness, and motion software issues | Customers were told they may continue using affected systems in accordance with their intended use. | 539 | Philips recalls Spectral CT systems over image, slice-thickness, and motion software issues — read full entry |
| QUANTUM OPS IncADAPTABLE Beach Chair surgical patient positioner, product numbers 804-33-300 and 3105-400-030 | Quantum Ops recalls ADAPTABLE Beach Chair over head-rest weld failure | Customers were instructed to locate affected devices and perform the specified safety check. | 266 | Quantum Ops recalls ADAPTABLE Beach Chair over head-rest weld failure — read full entry |
Entry 01
Micro Therapeutics recalls Pipeline Flex lot over incorrect implant diameter labeling
- Company/Product
- Micro Therapeutics, Inc. — Pipeline Flex Embolization Device with Shield Technology, REF PED2-300-12, lot D112860
- Recall issue
- Micro Therapeutics recalls Pipeline Flex lot over incorrect implant diameter labeling
- Official action
- Customers were instructed not to use affected devices and to remove and quarantine unused affected product.
- Quantity
- 2
Entry 02
Tandem recalls beta t:slim mobile app after Libre 3 Plus pairing failure
- Company/Product
- Tandem Diabetes Care, Inc. — Tandem t:slim mobile application, iOS software version 2.13 (TestFlight)
- Recall issue
- Tandem recalls beta t:slim mobile app after Libre 3 Plus pairing failure
- Official action
- Customers were advised to discontinue use of iOS t:slim mobile application version 2.13.
- Quantity
- 177
Entry 03
Spectrum Medical recalls Quantum PureFlow heat exchanger over blood-to-glycol leakage
- Company/Product
- SPECTRUM MEDICAL SRL — Quantum PureFlow Standard Heat Exchanger High Flow 3/8, model HX55W-S0
- Recall issue
- Spectrum Medical recalls Quantum PureFlow heat exchanger over blood-to-glycol leakage
- Official action
- Users were instructed to monitor the heat exchanger during procedures for a color change in the glycol compartment indicating leakage.
- Quantity
- 3232
Entry 04
Philips recalls Spectral CT 7500 over Bolus Tracking issue and potential additional radiation
- Company/Product
- Philips North America Llc — Spectral CT 7500, software version 5.2.x, product code 728333
- Recall issue
- Philips recalls Spectral CT 7500 over Bolus Tracking issue and potential additional radiation
- Official action
- Philips issued an Electronic Product Radiation Correction notification describing actions intended to prevent patient risk.
- Quantity
- 62
Entry 05
Siemens recalls Atellica CH RCRP assay over potential falsely elevated results
- Company/Product
- Siemens Healthcare Diagnostics, Inc. — Atellica CH Revised C-Reactive Protein (RCRP), Siemens Material Number 11537223, all lots
- Recall issue
- Siemens recalls Atellica CH RCRP assay over potential falsely elevated results
- Official action
- Customers were instructed to follow the supplied instructions to set up and calibrate the device.
- Quantity
- 42940 units (3938 US, 39002 OUS)
Entry 06
STERIS recalls VERIFY biological indicators and test packs over low spore population
- Company/Product
- Steris Corporation — STERIS VERIFY biological indicators and test packs
- Recall issue
- STERIS recalls VERIFY biological indicators and test packs over low spore population
- Official action
- Customers were instructed to destroy any affected product remaining on hand.
- Quantity
- LCB007: 409; S3060: 1185; S3061: 2205; S3065: 217; S3069: 233
Entry 07
Philips recalls Spectral CT systems over image, slice-thickness, and motion software issues
- Company/Product
- PHILIPS MEDICAL SYSTEMS — Philips Spectral CT and Spectral CT 7500 RT
- Recall issue
- Philips recalls Spectral CT systems over image, slice-thickness, and motion software issues
- Official action
- Customers were told they may continue using affected systems in accordance with their intended use.
- Quantity
- 539
Entry 08
Quantum Ops recalls ADAPTABLE Beach Chair over head-rest weld failure
- Company/Product
- QUANTUM OPS Inc — ADAPTABLE Beach Chair surgical patient positioner, product numbers 804-33-300 and 3105-400-030
- Recall issue
- Quantum Ops recalls ADAPTABLE Beach Chair over head-rest weld failure
- Official action
- Customers were instructed to locate affected devices and perform the specified safety check.
- Quantity
- 266
Fully expanded
This week’s entries
Official recall facts and actions are separated from MedTech Current significance analysis.
Micro Therapeutics recalls Pipeline Flex lot over incorrect implant diameter labeling
Pipeline Flex Embolization Device with Shield Technology, REF PED2-300-12, lot D112860
- Status
- Open, Classified
- Recalling firm
- Micro Therapeutics, Inc.
- Event ID
- 99689
- Recall number
- Z-3326-2026
- FDA posting date
- Recall initiated
- Quantity
- 2
- Distribution
- U.S. Nationwide distribution in the states of GA and TX.
- Identifiers
- REF PED2-300-12; Lot D112860
Official recall information
What happened
FDA classified the recall of Pipeline Flex Embolization Device with Shield Technology lot D112860 as Class II because devices labeled as 3.00 mm diameter contain 2.50 mm diameter implants. The product is a permanent intracranial aneurysm flow-diverter implant with a guidewire-based delivery system. FDA lists a Quantity in Commerce of 2, without a stated unit, distributed in Georgia and Texas.
MedTech Current analysis
Why it matters
The labeled diameter guides implant selection in a neurovascular procedure. A small reported quantity does not remove the concern: the implant remains permanently in the patient. The response therefore covers unused stock and patients already treated, whose records need review to assess whether follow-up is appropriate.
Official recall information
Official actions
According to the FDA recall record:
- Customers were instructed not to use affected devices and to remove and quarantine unused affected product.
- Affected product was to be returned to Medtronic, with replacement support available through the Medtronic representative.
- Customers were instructed to review records for patients treated with the affected lot and assess whether follow-up is appropriate.
- Customers were instructed to complete the confirmation form and share the recall communication with other affected organizations.
Tandem recalls beta t:slim mobile app after Libre 3 Plus pairing failure
Tandem t:slim mobile application, iOS software version 2.13 (TestFlight)
- Status
- Open, Classified
- Recalling firm
- Tandem Diabetes Care, Inc.
- Event ID
- 99589
- Recall number
- Z-3330-2026
- FDA posting date
- Recall initiated
- Quantity
- 177
- Distribution
- US Nationwide distribution.
- Identifiers
- iOS version 2.13 (TestFlight)
Official recall information
What happened
FDA classified Tandem's beta t:slim mobile application version 2.13 for iOS as Class II after a software issue prevented pairing with Abbott FreeStyle Libre 3 Plus sensors. Affected users may lose CGM values in the app and automated closed-loop operation, requiring manual pump use and potentially contributing to hypoglycemia or hyperglycemia. FDA lists a Quantity in Commerce of 177, without a stated unit.
MedTech Current analysis
Why it matters
Losing CGM connectivity changes how the insulin pump operates: users must manage delivery in manual mode without the app’s glucose values or automated control. For connected-device teams, the failure links release controls and sensor-compatibility testing to the ability to restore a working production version quickly.
Official recall information
Official actions
According to the FDA recall record:
- Customers were advised to discontinue use of iOS t:slim mobile application version 2.13.
- Customers were instructed to uninstall version 2.13 from their mobile device.
- Customers were instructed to return to production version 2.12.1 from the Apple App Store.
- Users were instructed to scan a compatible FreeStyle Libre 3 Plus sensor with version 2.12.1 when starting the next sensor session.
Spectrum Medical recalls Quantum PureFlow heat exchanger over blood-to-glycol leakage
Quantum PureFlow Standard Heat Exchanger High Flow 3/8, model HX55W-S0, lot 18051160301178
- Status
- Open, Classified
- Recalling firm
- SPECTRUM MEDICAL SRL
- Event ID
- 99462
- Recall number
- Z-3328-2026
- FDA posting date
- Recall initiated
- Quantity
- 3232
- Distribution
- US Nationwide distribution in the states of AL, CA, FL, KS, MA, MD, MI, MO, NC, NJ, NY, OH, OR, PA, SC, TN, TX, VA, and WI.
- Identifiers
- Model HX55W-S0; Lot 18051160301178
Official recall information
What happened
FDA classified the Quantum PureFlow Standard Heat Exchanger High Flow 3/8 recall as Class II following blood-to-glycol leakage between compartments during cardiopulmonary bypass. The single-use device heats or cools circulating blood through a compatible heater-cooler. The recall covers model HX55W-S0, lot 18051160301178. FDA lists 3232, without a quantity unit, and identifies process design as the cause.
MedTech Current analysis
Why it matters
A leak during bypass requires a response in the circuit and to the equipment supporting it. Replacing the disposable heat exchanger does not resolve contamination of the heater-cooler; that equipment also needs removal from clinical use, decontamination, and inspection.
Official recall information
Official actions
According to the FDA recall record:
- Users were instructed to monitor the heat exchanger during procedures for a color change in the glycol compartment indicating leakage.
- If leakage is observed, the heat exchanger should be immediately replaced in the circuit and returned to Spectrum Medical.
- Remove a contaminated heater-cooler from clinical use and return it for decontamination and inspection; do not use it with another patient.
- Customers using a third-party heater-cooler were instructed to remove it from clinical use and contact its manufacturer if blood contamination occurs.
Philips recalls Spectral CT 7500 over Bolus Tracking issue and potential additional radiation
Spectral CT 7500, software version 5.2.x, product code 728333
- Status
- Open, Classified
- Recalling firm
- Philips North America Llc
- Event ID
- 99865
- Recall number
- Z-3155-2026
- FDA posting date
- Recall initiated
- Quantity
- 62
- Distribution
- U.S. and U.S. Territories.
- Identifiers
- Product code 728333; Software version 5.2.x
Official recall information
What happened
FDA classified a Philips Spectral CT 7500 correction as Class II because the Bolus Tracking scan application may crash and Bolus Tracker shots may continue after the contrast threshold is reached. FDA states that this can result in additional radiation exposure to the patient. The affected configuration uses software version 5.2.x, and FDA lists a Quantity in Commerce of 62, without a stated unit.
MedTech Current analysis
Why it matters
The software failure can change radiation delivered during a scan, beyond disrupting system availability. Imaging departments must manage the affected workflow while the correction is arranged. For software-quality teams, it shows how failure to stop a scan sequence can become a radiation-control problem.
Official recall information
Official actions
According to the FDA recall record:
- Philips issued an Electronic Product Radiation Correction notification describing actions intended to prevent patient risk.
- Philips stated that it will schedule appointments with affected customers to install a software solution free of charge.
Siemens recalls Atellica CH RCRP assay over potential falsely elevated results
Atellica CH Revised C-Reactive Protein (RCRP), Siemens Material Number 11537223, all lots
- Status
- Open, Classified
- Recalling firm
- Siemens Healthcare Diagnostics, Inc.
- Event ID
- 99726
- Recall number
- Z-3335-2026
- FDA posting date
- Recall initiated
- Quantity
- 42940 units (3938 US, 39002 OUS)
- Distribution
- Worldwide distribution, including U.S. nationwide distribution and multiple international markets.
- Identifiers
- Siemens Material Number 11537223; All lots
Official recall information
What happened
FDA classified Siemens' Atellica CH Revised C-Reactive Protein assay recall as Class II because of the potential for falsely elevated results. The recall applies to all lots of Siemens Material Number 11537223. FDA reports 42940 units in commerce, including 3938 in the United States and 39002 outside the United States, with worldwide distribution and the cause still under investigation.
MedTech Current analysis
Why it matters
An operational analyzer can produce results that mislead clinical interpretation. Beyond calibration and discarded tests, the Medical Director must decide whether earlier results need action. Laboratory and IVD quality teams face an assay correction and a review of its clinical implications.
Official recall information
Official actions
According to the FDA recall record:
- Customers were instructed to follow the supplied instructions to set up and calibrate the device.
- Customers were instructed to record unused tests discarded for reimbursement or credit.
- Laboratories were instructed to review the letter with their Medical Director and determine the appropriate course of action, including for previously generated results if applicable.
- Return the Field Correction Effectiveness Check Form within 30 days, retain the letter with laboratory records, and forward it to recipients of affected product.
STERIS recalls VERIFY biological indicators and test packs over low spore population
VERIFY LCB007, S3065, and S3069: lots 260810, 260810A, 261106, 270116; VERIFY S3060 and S3061: those lots plus 270302. Model-specific details and expiration dates appear in the five linked FDA records.
- Status
- Open, Classified
- Recalling firm
- Steris Corporation
- Event ID
- 99704
- Recall number
- Z-0071-2027, Z-0072-2027, Z-0073-2027, Z-0074-2027, Z-0075-2027
- FDA posting date
- Recall initiated
- Quantity
- LCB007: 409; S3060: 1185; S3061: 2205; S3065: 217; S3069: 233
- Distribution
- Worldwide - US Nationwide distribution.
- Identifiers
- LCB007, S3065, S3069: lots 260810, 260810A, 261106, 270116; S3060, S3061: lots 260810, 260810A, 261106, 270116, 270302
Official recall information
What happened
FDA posted five Class II records under event 99704 for STERIS VERIFY biological indicators, test packs, and challenge packs. The Geobacillus stearothermophilus spore population may fall below the minimum shelf-life specification, potentially producing a no-growth result even when the steam cycle did not achieve the intended spore reduction. FDA reports record-level quantities of 409, 1185, 2205, 217, and 233 without stated units.
MedTech Current analysis
Why it matters
A no-growth result could give sterile-processing staff false reassurance that a steam cycle was effective. Affected stock must be identified and destroyed, while replacement depends on remaining shelf life. The five records belong together because they cover the same product family, failure, firm, and corrective instructions.
Official recall information
Official actions
According to the FDA recall records:
- Customers were instructed to destroy any affected product remaining on hand.
- STERIS stated that it will replace affected product that is still within its shelf life.
Philips recalls Spectral CT systems over image, slice-thickness, and motion software issues
Specified serial numbers in the FDA record: model 728333 in the U.S. and outside the U.S.; models 728340, 728344, and 728345 outside the U.S. Software 5.2.x in the U.S.; specified 5.0.x, 5.2.x, and 5.4.x versions outside the U.S.
- Status
- Open, Classified
- Recalling firm
- PHILIPS MEDICAL SYSTEMS
- Event ID
- 99796
- Recall number
- Z-0078-2027
- FDA posting date
- Recall initiated
- Quantity
- 539
- Distribution
- Worldwide - US Nationwide distribution and multiple international markets.
- Identifiers
- Models 728333, 728340, 728344, 728345; Affected serial numbers: see FDA Code Information
Official recall information
What happened
FDA posted a separate Class II Philips Spectral CT correction on October 2. Unlike the September 29 radiation-control recall, this event concerns software issues that may produce image artifacts, wrong slice thickness, or unintended system motion across Spectral CT and Spectral CT 7500 RT configurations. FDA lists a Quantity in Commerce of 539, without a stated unit, with worldwide distribution.
MedTech Current analysis
Why it matters
Image artifacts and wrong slice thickness can affect interpretation; unintended motion also concerns system behavior. Radiology, clinical engineering, and software change-control teams need the instructions for this event. It remains separate from the Bolus Tracking recall because its failure modes and corrective instructions differ.
Official recall information
Official actions
According to the FDA recall record:
- Customers were told they may continue using affected systems in accordance with their intended use.
- Facilities were instructed to circulate the notice to all users so they are aware of the issues, potential hazards, and manufacturer recommendations.
- Facilities were instructed to post the notice near affected units until a solution is installed.
- Customers were instructed to complete and return the response form within 30 days.
Quantum Ops recalls ADAPTABLE Beach Chair over head-rest weld failure
ADAPTABLE Beach Chair surgical patient positioner, product numbers 804-33-300 and 3105-400-030; only affected lots and serial numbers listed in FDA Code Information, with the stated demo/undelivered exclusions
- Status
- Open, Classified
- Recalling firm
- QUANTUM OPS Inc
- Event ID
- 99384
- Recall number
- Z-0076-2027
- FDA posting date
- Recall initiated
- Quantity
- 266
- Distribution
- U.S. Nationwide distribution in the states of CA and TX.
- Identifiers
- Product numbers 804-33-300 and 3105-400-030; UDI-DI 00190446667278; Affected lots and serial numbers: see FDA Code Information
Official recall information
What happened
FDA classified the ADAPTABLE Beach Chair surgical patient positioner recall as Class II because an inadequate weld in the patient head-positioning system can fracture below the head rest. The firm received complaints describing welded-joint fracture and head-rest separation. FDA lists a Quantity in Commerce of 266, without a stated unit, distributed in California and Texas, and identifies device design as the determined cause.
MedTech Current analysis
Why it matters
The positioner supports the patient’s upper body and head during surgery; a failed joint can disrupt positioning. Facilities must locate and check affected devices. Continued use depends on passing the safety check. Passing devices still need exchange for an upgraded unit.
Official recall information
Official actions
According to the FDA recall record:
- Customers were instructed to locate affected devices and perform the specified safety check.
- Devices that pass the safety check may remain in clinical use pending exchange for an upgraded unit.
- Devices that fail the safety check must be segregated and removed from use, with return and replacement arranged.
- Notify staff and organizations that received the product. Maintain awareness until actions are complete; return the acknowledgement or reply form even if no affected product is found.
This issue retains the supplied 28 September–4 October 2026 window and includes records by FDA Recall Posting Date. Eight selected events are represented by 12 linked FDA records; the five STERIS records are grouped by firm, product family, failure, and corrective action. Quantities without stated units are preserved without added units or aggregation.